![Emerson Rosemount OCX 8800 Скачать руководство пользователя страница 130](http://html.mh-extra.com/html/emerson/rosemount-ocx-8800/rosemount-ocx-8800_instruction-manual_104777130.webp)
Section 7: Foundation
Fieldbus
April 2017
Instruction Manual
IM-106-880C, OI
Parameter Mnemonic Valid Range Units
Description
Table 7-4 Transducer Block Parameter Description
ALARM_RELAY_EVENT1
Enumerated The first of three conditions that cause the alarm output to turn on..
ALARM_RELAY_EVENT2
Enumerated The second of three conditions that cause the alarm output to turn
on.
ALARM_RELAY_EVENT3
Enumerated The third of three conditions that cause the alarm output to turn on.
ALARM_RELAY_STATE
0: Off – 1: On
Enumerated The state of the alarm output.
ALERT_KEY
See FF-891 section 5.3.
ANALYZER_SW_BUILD_
DATE
ANALYZER_SW_BUILD_
NUMBER
ANALYZER_SW_
CHECKSUM
The date the analyzer software was built.
The build number of the analyzer software.
The checksum of the analyzer software.
ANALYZER_SW_VERSION
The version of the analyzer software.
BLOCK_ALM
See FF-891 section 5.3.
BLOCK_ERR
See FF-891 section 5.3.
BLOWBACK_DURATION
1-5
Seconds
The amount of time the blowback solenoid will be on.
BLOWBACK_ENABLED
0: No – 1: Yes
Enumerated Enables or disables the automatic blowback cycle.
BLOWBACK_INTERVAL
0-32767
Minutes
The time between blowback cycles.
BLOWBACK_PURGE_TIME 0-500
Seconds
The amount of time before returning the output to process after
performing a blowback.
BLOWBACK_STATE
0, 1, 2
Enumerated The current state of the blowback cycle. (0=Idle, 1=Blow, 2=Purge)
CAL_GAS_TIME
60-1200
Seconds
The amount of time calibration gas should flow before a reading is
taken.
CAL_PURGE_TIME
60-1200
Seconds
The amount of time before returning the output to process after
calibrating.
CAL_REC_ENABLE
0: No – 1: Yes
Enable/disable calibration recommended alarm.
CAL_RESULTS
Bit Enum
Calibration result.
CAL_STATE
Enumerated The current state of the calibration cycle.
CAL_STATE_STEP
Initiates a calibration or goes to the next calibration step.
CAL_STATE_TIME
Seconds
The time left in the current calibration step.
COLLECTION_DIRECTORY
See Transducer Block Specification, part 1. FF-902, page 11.
COMB _SENSOR_CAL_LOC
See FF-903 section 3.3.
COMB_SENSOR_CAL_
METHOD
COMB _SENSOR_-
CAL_WHO
COMB_AUTOCAL_
INTERVAL
See FF-903 sections 3.3 and 4.5.
See FF-903 section 3.3.
0-9999
Hours
The time between automatic calibrations of the combustibles sen-
sor.
COMB_CAL_POINT
0-55000
PPM
The value of the combustibles test gas.
COMB_CONSTANT
-99.0-99.0
PPM
The combustibles calibration constant.
COMB_DELTA_
RESISTANCE
Ohms
The raw value of the combustibles level input. This is the difference
between the active and reference RTDs.
COMB_FAILED_CONSTANT
PPM
This is the constant value calculated from the last failed
Combustibles calibration.
COMB_FAILED_SLOPE
PPM/Ohm
This is the slope value calculated from the last failed Combustibles
calibration.
COMB_HTR_DUTYCYCLE
Combustibles heater duty cycle.
104
Transducer Block Parameters
Содержание Rosemount OCX 8800
Страница 2: ......
Страница 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Страница 192: ...Section 7 Foundation Fieldbus April 2017 Instruction Manual IM 106 880C OI 16 6 About AMS and DeltaV Software ...
Страница 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Страница 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Страница 259: ...Section 10 Replacement Parts April 2017 Instruction Manual IM 106 880C OI 23 2 O2 Cell and Heater Strut Assembly ...
Страница 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Страница 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Страница 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Страница 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Страница 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Страница 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Страница 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Страница 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Страница 298: ...SPA with HART Alarm April 2017 Instruction Manual IM 106 880C OI 27 2 Configuration Calibratio n ...
Страница 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Страница 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Страница 305: ......