Instruction Manual
IM-106-880C, OI
Section 5:
Calibration
April 2017
5.6
D/A Trim Procedures - LOI
5.6.1
O
2
D/A trim procedure using the LOI
Use the following procedure to perform the O
2
D/A trim procedure at the LOI. Refer to the LOI
menu tree in
Figure 4-5
.
To avoid a potentially dangerous operating condition, the Hazardous Area OCX 8800
must be removed from the automatic combustion control loop before you start the D/A
trim procedure.
1.
From the operating display use the left-pointing key to select the first column submenu.
Use the down-pointing key to select SYSTEM.
2.
From the SYSTEM menu, use the down-pointing key to select Input/Output. Use the
right-pointing key to select the Analog parameters list.
3.
Scroll down to the item Trim O2 Out. Touch the right-pointing key to start the O
2
trim
procedure.
NOTE
If you wish to exit D/A Trim with no changes, step through the procedure using yes responses,
and enter no meter readings.
4.
Remove the electronics housing cover.
5.
Refer to Figure 2-6. Connect a digital multimeter to read the milliamp output from the
O
2
D/A converter circuit. Connect the positive lead to the AOUT1+ terminal and
connect the negative lead to the AOUT1- terminal. Then, touch the Enter key at the LOI.
6.
The LOI displays 4 mA........Meter. The trim program inputs the design-equivalent signal
for a 4.00 mA output. Read the O
2
milliamp output at the digital multimeter. Use the
right-pointing key to select each digit and use the up- and down-pointing keys to
change the value. When the correct value is displayed, use the Enter key to input the
value.
7.
The LOI displays 20 mA........Meter. The trim program inputs the design-equivalent
signal for a 20.00 mA output. Read the O
2
milliamp output at the digital multimeter.
Use the right-pointing key to select each digit and use the up- and down-pointing keys
to change the value. When the correct value is displayed, use the Enter key to input the
value.
8.
The LOI displays a Meter at 4 mA prompt. Use the right-pointing key to select the letter
yes or no. Use the up- or down-pointing key to change the letter. Then use the Enter key
to input the response. If no, the process repeats from step 7.
9.
The LOI displays a Meter at 20 mA prompt. Use the right-pointing key to select the letter
yes or no. Use the up- or down-pointing key to change the letter. Then use the Enter key
to input the response. If no, the process repeats from step 8.
10.
When the ropiness in steps 8 and 9 are yes, the trim procedure is complete. Exit the LOI
menu and return the control loop to automatic control.
D/A Trim Procedures - LOI
75
Содержание Rosemount OCX 8800
Страница 2: ......
Страница 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Страница 192: ...Section 7 Foundation Fieldbus April 2017 Instruction Manual IM 106 880C OI 16 6 About AMS and DeltaV Software ...
Страница 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Страница 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Страница 259: ...Section 10 Replacement Parts April 2017 Instruction Manual IM 106 880C OI 23 2 O2 Cell and Heater Strut Assembly ...
Страница 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Страница 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Страница 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Страница 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Страница 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Страница 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Страница 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Страница 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Страница 298: ...SPA with HART Alarm April 2017 Instruction Manual IM 106 880C OI 27 2 Configuration Calibratio n ...
Страница 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Страница 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Страница 305: ......