8.2
Chlorine sensor
8.2.1
General
When used in clean water, the sensor requires little maintenance. Generally, the sensor
needs maintenance when the response becomes sluggish or noisy or when readings drift
following calibration.
Maintenance frequency is best determined by experience. For a sensor used in potable
water, expect to clean the membrane every month and replace the membrane and
electrolyte solution every three months.In water containing large amounts of suspended
solids, for example, open recirculating cooling water, membrane cleaning or replacement
will be more frequent.
8.2.2
Cleaning the membrane
Clean the membrane with water sprayed from a wash bottle.
Important
Do not use tissues to clean the membrane.
8.2.3
Replacing the electrolyte solution and membrane
WARNING
HARMFUL SUBSTANCE
Fill solution may cause irritation. May be harmful if swallowed. Read and follow manual.
Procedure
1. Unscrew the membrane retainer.
2. Remove the membrane assembly and O-ring.
See
.
Maintenance
Reference Manual
May 2019
00809-0100-3412
78
Emerson.com/Rosemount
Содержание Rosemount FCL
Страница 4: ...4 ...
Страница 10: ...Description and specifications Reference Manual May 2019 00809 0100 3412 10 Emerson com Rosemount ...
Страница 18: ...Install Reference Manual May 2019 00809 0100 3412 18 Emerson com Rosemount ...
Страница 26: ...Wire Reference Manual May 2019 00809 0100 3412 26 Emerson com Rosemount ...
Страница 27: ...Reference Manual Wire 00809 0100 3412 May 2019 Rosemount FCL 1056 27 ...
Страница 28: ...Wire Reference Manual May 2019 00809 0100 3412 28 Emerson com Rosemount ...
Страница 76: ...Digital communications Reference Manual May 2019 00809 0100 3412 76 Emerson com Rosemount ...
Страница 107: ...Reference Manual 00809 0100 3412 May 2019 Rosemount FCL 1056 107 ...