Troubleshooting and diagnostics
7.1
Troubleshooting, repairs, and failure
diagnostics
The analyzer is specifically designed to run unattended for long periods, to automatically
resolve system issues, and to recover from power failures and return to a normal working
state without intervention.
This troubleshooting guide is intended to assist maintenance personnel when the analyzer
has not appeared to be working normally for a period of more than five minutes. If the
procedures in this chapter fail to return the analyzer to normal operations or do not
identify a fault, notify your Rosemount Customer Care Representative
(
) at which point the analyzer may require to be returned
for detailed examination and fault finding at a service center.
Observe and obey all safety precautions when performing preventative maintenance on
the analyzer.
Important
The troubleshooting and failure diagnosis procedures described in this chapter assume
that any host equipment provided by you is fully functional. Always make sure that the
host equipment is fully serviceable before performing failure diagnosis on the analyzer.
Failure diagnosis, repair, and maintenance must only be performed by:
•
Maintenance engineers who have the necessary skills, and who have been authorized
to perform maintenance on the analyzer.
•
Emerson customer care engineers.
In all the cases described in this chapter, maintenance personnel must perform the repairs
by directly replacing the faulty item with a known serviceable spare part supplied by
Emerson. All other items must be repaired or replaced by the manufacturer.
CAUTION
EQUIPMENT DAMAGE
In the event of a power failure any residual sample gas should be purged from the analyzer
to ensure any corrosive gas elements are not left to contaminate/degrade the internal
components. Failure to do so could result in irreparable damage to the analyzer.
Reference Manual
Troubleshooting and diagnostics
00809-0100-4244
January 2020
Reference Manual
65
Содержание Rosemount CT4400
Страница 1: ...Reference Manual 00809 0100 4244 Rev AA January 2020 Rosemount CT4400 Continuous Gas Analyzer ...
Страница 32: ...Start up procedure Reference Manual January 2020 00809 0100 4244 32 Rosemount CT4400 ...
Страница 40: ...Operating the analyzer Reference Manual January 2020 00809 0100 4244 40 Rosemount CT4400 ...
Страница 64: ...Gas calibration Reference Manual January 2020 00809 0100 4244 64 Rosemount CT4400 ...
Страница 96: ...Shutdown procedure Reference Manual January 2020 00809 0100 4244 96 Rosemount CT4400 ...
Страница 98: ...Preventative maintenance Reference Manual January 2020 00809 0100 4244 98 Rosemount CT4400 ...
Страница 100: ...Recommended spare parts Reference Manual January 2020 00809 0100 4244 100 Rosemount CT4400 ...
Страница 105: ......