Gas calibration
6.1
Required tools
The gas concentrations measured by the analyzer can be validated against a known
sample gas or calibrated to match it by using the following gas calibration procedure.
To calibrate the analyzer, you need the following items:
•
Nitrogen gas of instrument gas purity for use as a zero calibration gas
•
Suitable span calibration gases for each gas measured
•
Gas bottle pressure regulators
•
Interconnecting hoses to connect the gas bottles to the analyzer
•
A T-piece and excess flow line
NOTICE
In the case of gases, such as H
2
O, for which it is not normally possible to obtain calibrated
gas cylinders, it will usually be measured by the same laser as some other gas. Validating
the other gases measured by the analyzer (particularly any which are measured by the
same laser as H
2
O) can demonstrate that the system is functioning correctly, meaning
there is no need to calibrate the water measurement directly.
If calibration must be carried out (e.g., for legal requirements), it can be calibrated by
using a water vapor generator to supply a known concentration of water vapor.
6.2
Gas service menu definitions
The
GAS SERVICE
menu screen can be accessed from the
Main Menu
screen.
This gives the operator access to a number of software routines and screens that are used
for calibration and maintenance of the analyzer.
Where more options are available than can be displayed on the screen use the
or
keys to scroll up or down onto the next page. The screen will continually wrap around until
the option required is highlighted.
Pressing the
Forward key
will move to the next screen where available as shown in the
GAS Service/Per Gas
example below.
Pressing the screen
Refresh key
at any time will refresh the current display screen if
required by the operator.
PER GAS
allows for gas calibration or verification on a gas by gas basis.
CALIBR ALL
not an option available for the Rosemount CT4400 analyzer.
VERIFY ALL
not an option available for the Rosemount CT4400 analyzer.
Reference Manual
Gas calibration
00809-0100-4244
January 2020
Reference Manual
49
Содержание Rosemount CT4400
Страница 1: ...Reference Manual 00809 0100 4244 Rev AA January 2020 Rosemount CT4400 Continuous Gas Analyzer ...
Страница 32: ...Start up procedure Reference Manual January 2020 00809 0100 4244 32 Rosemount CT4400 ...
Страница 40: ...Operating the analyzer Reference Manual January 2020 00809 0100 4244 40 Rosemount CT4400 ...
Страница 64: ...Gas calibration Reference Manual January 2020 00809 0100 4244 64 Rosemount CT4400 ...
Страница 96: ...Shutdown procedure Reference Manual January 2020 00809 0100 4244 96 Rosemount CT4400 ...
Страница 98: ...Preventative maintenance Reference Manual January 2020 00809 0100 4244 98 Rosemount CT4400 ...
Страница 100: ...Recommended spare parts Reference Manual January 2020 00809 0100 4244 100 Rosemount CT4400 ...
Страница 105: ......