3.3
Start-up procedure
NOTICE
The gases shown in the screenshots and the measurements may be different from those
shown in your particular analyzer. They indicate the functionality of the software, which is
the same regardless of the gases being measured.
CAUTION
To stop the start-up procedure at any time, set the main circuit breaker to
OFF
.
To start the analyzer, perform the following steps:
Procedure
1. Visually inspect the exterior of the analyzer for signs of damage, corrosion, gas
leaks, or overheating. Report anything found to the maintenance organization.
2. Ensure that the analyzer has been correctly installed.
See
3. Ensure that the
On/Off
switch at the rear of the analyzer is set to
I
(
ON
).
Figure 3-1: Rear Panel ON/OFF Switch
A. Rear panel
B. On/Off switch
4. The start-up sequence commences automatically under software control.
5. After a few minutes, the
Gas Sensor Main
) appears on the display
controller. If it does not, report the fault to maintenance.
Start-up procedure
Reference Manual
January 2020
00809-0100-4244
30
Rosemount CT4400
Содержание Rosemount CT4400
Страница 1: ...Reference Manual 00809 0100 4244 Rev AA January 2020 Rosemount CT4400 Continuous Gas Analyzer ...
Страница 32: ...Start up procedure Reference Manual January 2020 00809 0100 4244 32 Rosemount CT4400 ...
Страница 40: ...Operating the analyzer Reference Manual January 2020 00809 0100 4244 40 Rosemount CT4400 ...
Страница 64: ...Gas calibration Reference Manual January 2020 00809 0100 4244 64 Rosemount CT4400 ...
Страница 96: ...Shutdown procedure Reference Manual January 2020 00809 0100 4244 96 Rosemount CT4400 ...
Страница 98: ...Preventative maintenance Reference Manual January 2020 00809 0100 4244 98 Rosemount CT4400 ...
Страница 100: ...Recommended spare parts Reference Manual January 2020 00809 0100 4244 100 Rosemount CT4400 ...
Страница 105: ......