Note
Zero mA should be set up no matter what has been selected as the base of the current output. This is
factory set.
In a similar manner to the above, set the current output level to 20 mA.
3.7.4
Parameters
Complete the following steps to set the parameters.
1.
With this option displayed, press ENTER to access the list of six available options.
The arrow keys cycle through these options.
2.
When the option you need to change is displayed, press ENTER.
3.
When you have made all required changes, select
EXIT
and press ENTER.
The six available options are:
a. Security Number
b. Identity Number
c. Path Length
d. Alarm
e. Cal. Factor
f. Plant Status Input
Security Number
To prevent any unauthorized tampering with the setup information, it is important to
change the security code from the factory setting.
Select each digit with ENTER and change it with the arrow keys.
Note
It is important to make note of this number; otherwise, it will not be possible to change the
instrument set up.
Configuration and startup
40
Rosemount CCO 5500
Содержание Rosemount CCO 5500
Страница 1: ...Reference Manual 00809 0100 4550 Rev AA June 2018 Rosemount CCO 5500 Carbon Monoxide CO Analyzer ...
Страница 4: ......
Страница 8: ...Contents iv Rosemount CCO 5500 ...
Страница 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Страница 32: ...Install 24 Rosemount CCO 5500 ...
Страница 88: ...Operation 80 Rosemount CCO 5500 ...
Страница 98: ...Maintenance 90 Rosemount CCO 5500 ...
Страница 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Страница 110: ...Spare parts 102 Rosemount CCO 5500 ...
Страница 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Страница 123: ...Return equipment to the factory Reference Manual 115 ...