The communication protocol is Modbus
®
compatible.
—
This protocol is standard and widely used.
—
It enables continuous communication between a standard Modbus
controller (master device) and a serial network of up to 247 detectors.
Heated
optics
The front window can be heated to improve performance in icing,
condensation, and snow conditions. The heater increases the temperature
of the optical surface by 5 to 8 °F (3 to 5 °C) above the ambient temperature.
The heated optics can be configured in three ways:
•
OFF: The optics are not heated.
•
ON: The optics are heated continuously.
•
AUTO: Operated only when the change of temperature requires heating
(default).
In AUTO mode, you can define the start heating temperature between 32
and 122 °F (0 and 50 °C). The detector stops heating the window when the
temperature is 27 °F (15 °C) above the start temperature.
A.4
Mechanical specifications
Enclosure
Stainless steel 316
Water and dust tight
•
NEMA
®
250 type 6p.
•
IP 66 and IP 67 per EN 60529
Electronic modules
Conformal coated
Electrical connection (two entries)
•
¾
-in. -14 NPT conduit or
•
M25 x 1.5
Dimensions
4 x 4.6 x 6.18 in. (101.6 x 117 x 157 mm)
Weight
6.1 lb. (2.8 kg)
A.5
Environmental specifications
The Rosemount 975MR is designed to withstand harsh environmental conditions.
High
temperature
•
Designed to meet MIL-STD-810C, method 501.1, procedure II
•
Operating temperature: 167 °F (75 °C)
•
Storage temperature: 185 °F (85 °C)
Low
temperature
•
Designed to meet MIL-STD-810C, method 502.1, procedure I
•
Operating temperature: -57 °F (-50 °C)
•
Storage temperature: -65 °F (-55 °C)
Specifications and reference data
Reference Manual
August 2019
00809-0100-4975
50
Emerson.com/Rosemount
Содержание Rosemount 975MR
Страница 38: ...Installing the detector Reference Manual August 2019 00809 0100 4975 38 Emerson com Rosemount ...
Страница 46: ...Maintenance and troubleshooting Reference Manual August 2019 00809 0100 4975 46 Emerson com Rosemount ...
Страница 52: ...Specifications and reference data Reference Manual August 2019 00809 0100 4975 52 Emerson com Rosemount ...
Страница 60: ...RS 485 communication network Reference Manual August 2019 00809 0100 4975 60 Emerson com Rosemount ...
Страница 67: ...Figure D 4 Duct Mount Reference Manual Accessories 00809 0100 4975 August 2019 Rosemount 975MR 67 ...
Страница 70: ...Accessories Reference Manual August 2019 00809 0100 4975 70 Emerson com Rosemount ...