Table A-2: Electrical Specifications
(continued)
Operating voltage
Status
All outputs
Without 0-20 mA
Power consumption
(max. 18-32 Vdc)
Normal
1.95 W
1.85 W
Normal when heater
on
2.56 W
2.45 W
Alarm
3.04 W
2.56 W
Alarm when heater on
3.68 W
3.2 W
Maximum current (18 -
32 Vdc)
Normal
90 mA
85 mA
Normal when heater
on
105 mA
100 mA
Alarm
130 mA
115 mA
Alarm when heater on
160 mA
145 mA
Electrical input protection: The input circuit is protected against voltage-reversed polarity,
voltage transients, surges, and spikes according to MIL-STD-1275B.
A.3
Outputs
Electrical
interface
There are five output configurations. These configurations must be defined
at the factory per the customer order and cannot be changed at the
customer facility.
See
General instructions for electrical wiring
diagrams for each option.
Unless otherwise specified, the default is output configuration 1A. The
wiring arrangement is identified on the detector by the part number (see
).
•
Output configuration 1A: power, RS-485, 0-20 mA (sink), fault relay
(N.C.), alarm rela (N.O.)
•
Output configuration 2A: power, RS-485, 0-20 mA (source) and HART
®
protocol, fault relay (N.O.), alarm relay (N.O., N.C.)
•
Output configuration 3A: power, RS-485, 0-20 mA (source) and HART
protocol, fault relay (N.O.), alarm relay (N.O., N.C.)
•
Output configuration 1R: power, RS-485, fault relay (N.C.), auxiliary relay
(N.O.), alarm relay (N.O.)
•
Output configuration 2R: power, RS-485, fault relay (N.O.), auxiliary relay
(N.O.), alarm relay (N.O.)
Electrical
outputs
•
Dry contact relays
Table A-3: Contact Ratings
Relay name
Type
Normal position
Maximum
ratings
Alarm
SPDT
N.O., N.C.
2 A at 30 Vdc
Specifications and reference data
Reference Manual
August 2019
00809-0100-4975
48
Emerson.com/Rosemount
Содержание Rosemount 975MR
Страница 38: ...Installing the detector Reference Manual August 2019 00809 0100 4975 38 Emerson com Rosemount ...
Страница 46: ...Maintenance and troubleshooting Reference Manual August 2019 00809 0100 4975 46 Emerson com Rosemount ...
Страница 52: ...Specifications and reference data Reference Manual August 2019 00809 0100 4975 52 Emerson com Rosemount ...
Страница 60: ...RS 485 communication network Reference Manual August 2019 00809 0100 4975 60 Emerson com Rosemount ...
Страница 67: ...Figure D 4 Duct Mount Reference Manual Accessories 00809 0100 4975 August 2019 Rosemount 975MR 67 ...
Страница 70: ...Accessories Reference Manual August 2019 00809 0100 4975 70 Emerson com Rosemount ...