Advanced Diagnostics Configuration
Continuous limit
Set the test criteria for the
continuous SMART Meter Verification
diagnostic. The factory
default for this value is set to five percent with limits configurable between two and ten
percent. If the tolerance band is set too tightly, under empty pipe conditions or noisy
flowing conditions, a false failure of the transmitter test may occur.
6.10
Run manual SMART Meter Verification
The
SMART Meter Verification
diagnostic will be available if the advanced diagnostic suite
(DA2) was ordered. If DA2 was not ordered or licensed, this diagnostic will not be available.
This method will initiate the manual meter verification test.
6.10.1
Test conditions
SMART Meter Verification
can be initiated under three possible test conditions. This
parameter is set at the time that the
sensor baseline
or
SMART Meter Verification
test is
manually initiated.
No flow
Run the
SMART Meter Verification
test with a full pipe and no flow in the line. Running the
SMART Meter Verification
test under this condition provides the most accurate results and
the best indication of magnetic flowmeter health.
Flowing full
Run the
SMART Meter Verification
test with a full pipe and flow in the line. Running the
SMART
Meter Verification
test under this condition provides the ability to verify the magnetic
flowmeter health without shutting down the process flow in applications when a shutdown
is not possible. Running the diagnostic under flowing conditions can cause a false test
failure if there is significant process noise present.
Empty pipe
Run the
SMART Meter Verification
test with an empty pipe. Running the
SMART Meter
Verification
test under this condition provides the ability to verify the magnetic flowmeter
health with an empty pipe. Running the verification diagnostic under empty pipe conditions
will not check the electrode circuit health.
LOI menu path
Field Mount: Diagnostics, Advanced Diag, Meter Verify, Test Criteria,
Continual
Wall Mount: N/A
Traditional Fast Keys
1, 2, 3, 2, 4, 4
Device dashboard
2, 2, 6, 4, 1
LOI menu path
Field Mount: Diagnostics, Advanced Diag, Meter Verify, Run Meter Ver
Wall Mount: AUX. FUNCTION
Traditional Fast Keys
1, 2, 3, 2, 1
Device dashboard
1, 6
Содержание Rosemount 8750W
Страница 2: ......
Страница 96: ...84 Reference Manual 00809 0300 4750 Rev CA Operation March 2016 Operation ...
Страница 158: ...146 Reference Manual 00809 0300 4750 Rev CA Digital Signal Processing March 2016 Digital Signal Processing ...
Страница 168: ...156 Reference Manual 00809 0300 4750 Rev CA Maintenance March 2016 Maintenance ...
Страница 208: ...Product Certifications March 2016 Reference Manual 00809 0300 4750 Rev CA 196 Product Certifications ...
Страница 209: ...197 Product Certifications Product Certifications March 2016 Reference Manual 00809 0300 4750 Rev CA ...
Страница 212: ...Product Certifications March 2016 Reference Manual 00809 0300 4750 Rev CA 200 Product Certifications ...
Страница 214: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 202 Wiring Diagrams ...
Страница 218: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 206 Wiring Diagrams ...
Страница 219: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 207 Wiring Diagrams ...
Страница 220: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 208 Wiring Diagrams ...
Страница 222: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 210 Wiring Diagrams ...
Страница 223: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 211 Wiring Diagrams ...
Страница 224: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 212 Wiring Diagrams ...
Страница 225: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 213 Wiring Diagrams ...
Страница 227: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 215 Wiring Diagrams ...
Страница 228: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 216 Wiring Diagrams ...
Страница 229: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 217 Wiring Diagrams ...
Страница 263: ......