126
Reference Manual
00809-0300-4750, Rev CA
Advanced Diagnostics Configuration
March 2016
Advanced Diagnostics Configuration
Electrode circuit resistance
The
electrode circuit
resistance is a measurement of the electrode circuit health. This value is
used as a baseline to determine if the electrode circuit is still operating correctly.
6.9.2
Establishing the sensor baseline (signature)
The first step in running the
SMART Meter Verification
test is establishing the reference
signature that the test will use as the baseline for comparison. This is accomplished by
having the transmitter take a signature of the sensor.
Reset baseline (re-signature meter
)
Having the transmitter take an initial sensor signature when first installed will provide the
baseline for the verification tests that are done in the future. The sensor signature should be
taken during the start-up process when the transmitter is first connected to the sensor, with
a full line, and ideally with no flow in the line. Running the sensor signature procedure when
there is flow in the line is permissible, but this may introduce some noise into the
electrode
circuit resistance
measurement. If an empty pipe condition exists, then the sensor signature
should only be run for the coils.
Once the sensor signature process is complete, the measurements taken during this
procedure are stored in non-volatile memory to prevent loss in the event of a power
interruption to the meter. This initial sensor signature is required for both manual and
continuous SMART Meter Verification.
Recall values (recall last saved)
In the event that the sensor baseline was reset accidentally or incorrectly, this function will
restore the previously saved sensor baseline values.
LOI menu path
Field Mount: Diagnostics, Advanced Diag, Meter Verify, Sensr Baseline,
Values, Electrode Res
Wall Mount: XMTR INFO
Traditional Fast Keys
1, 2, 3, 2, 3, 1, 3
Device dashboard
2, 2, 6, 1, 3
LOI menu path
Field Mount: Diagnostics, Advanced Diag, Meter Verify, Sensr Baseline,
Reset Baseline
Wall Mount: AUX. FUNCTION
Traditional Fast Keys
1, 2, 3, 2, 3, 2
Device dashboard
2, 2, 6, 1, 4
LOI menu path
Field Mount: Diagnostics, Advanced Diag, Meter Verify, Sensr Baseline,
Recall Values
Wall Mount: AUX. FUNCTION
Traditional Fast Keys
1, 2, 3, 2, 3, 3
Device dashboard
2, 2, 6, 1, 5
Содержание Rosemount 8750W
Страница 2: ......
Страница 96: ...84 Reference Manual 00809 0300 4750 Rev CA Operation March 2016 Operation ...
Страница 158: ...146 Reference Manual 00809 0300 4750 Rev CA Digital Signal Processing March 2016 Digital Signal Processing ...
Страница 168: ...156 Reference Manual 00809 0300 4750 Rev CA Maintenance March 2016 Maintenance ...
Страница 208: ...Product Certifications March 2016 Reference Manual 00809 0300 4750 Rev CA 196 Product Certifications ...
Страница 209: ...197 Product Certifications Product Certifications March 2016 Reference Manual 00809 0300 4750 Rev CA ...
Страница 212: ...Product Certifications March 2016 Reference Manual 00809 0300 4750 Rev CA 200 Product Certifications ...
Страница 214: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 202 Wiring Diagrams ...
Страница 218: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 206 Wiring Diagrams ...
Страница 219: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 207 Wiring Diagrams ...
Страница 220: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 208 Wiring Diagrams ...
Страница 222: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 210 Wiring Diagrams ...
Страница 223: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 211 Wiring Diagrams ...
Страница 224: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 212 Wiring Diagrams ...
Страница 225: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 213 Wiring Diagrams ...
Страница 227: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 215 Wiring Diagrams ...
Страница 228: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 216 Wiring Diagrams ...
Страница 229: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 217 Wiring Diagrams ...
Страница 263: ......