Socket Module—Integral Mount
Figure 11-6:
Removing an integral mount socket module
1.
Disconnect power.
2.
Remove electronics cover to gain access to the coil and electrode cables.
3.
If the transmitter has an LOI, it will need to be removed to gain access to the coil and
electrode cables.
4.
Disconnect the coil and electrode cables.
5.
Remove the four transmitter mounting screws.
6.
Lift the transmitter off of the sensor adapter.
7.
To remove the socket module, loosen the two mounting screws and pull up on the
socket module from the base.
8.
When removing the socket module, do not pull on the wires.
.
Installing an integral mount socket module
1.
To insert a new integral mount socket module, press the base into its keyed position
and tighten the two mounting screws.
2.
The coil and electrode cables are fed through the bottom opening of the transmitter
and plugged into the face of the electronics.
3.
The coil and electrode cables are keyed so they will only fit into their dedicated
location.
4.
If the transmitter has an LOI, it will need to be removed to access the coil and
electrode ports.
5.
Once the connections are made, the transmitter can be secured to the sensor
adapter using the four mounting bolts.
Maintenance
146
Rosemount
®
8732EM Transmitter with HART Protocol Reference Manual
Содержание Rosemount 8732EM
Страница 2: ......
Страница 6: ...Contents iv Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 10: ...Safety messages 4 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 14: ...Introduction 8 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 30: ...Sensor Installation 24 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 87: ...Field Communicator Dashboard Menu Tree HART v5 4 part 2 Figure 7 8 Operation Reference manual 81 ...
Страница 92: ...Operation 86 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 138: ...Advanced Diagnostics Configuration 132 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 158: ...Maintenance 152 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 214: ...Product Certifications 208 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 217: ...Wiring 8732EM using combination cable Figure C 2 Wiring Diagrams Reference manual 211 ...
Страница 221: ...Wiring Diagram 475 Field Communicator with 8732EM External Analog Power Figure C 6 Wiring Diagrams Reference manual 215 ...
Страница 222: ...Wiring Diagrams 216 Rosemount 8732EM Transmitter with HART Protocol Reference Manual ...
Страница 253: ...Implementing a Universal Transmitter Reference manual 247 ...