Troubleshooting
00809-0100-4860, Rev BF
5.6.5
Remote electronics procedure
If the Rosemount 8600D electronics housing is mounted remotely, some replacement
procedures are different than for the flowmeter with integral electronics. The following
procedures are exactly the same:
Replacing the Terminal Block in the Housing (see
).
Replacing the Electronics Boards (see
).
Replacing the Sensor (see
).
To disconnect the coaxial cable from the meter body and electronics housing, follow the
instructions below.
Disconnect the coaxial cable at the meter
1.
Remove the access cover on the meter body bracket if present.
2.
Loosen the three housing rotation screws at the base of the meter adapter with a
5
/
32
-in. hex wrench by turning the screws clockwise (inward) until they clear the
bracket.
3.
Slowly pull the meter adapter no more than 1.5-in. (40 mm) from the top of the
bracket.
4.
Loosen and disconnect the sensor cable nut from the union using
a
5
/
16
-in. open end wrench.
Note
Do not pull the adapter more than 1.5-in. (40 mm) from the top of the bracket. Damage to
the sensor may occur if the sensor cable is stressed.
Содержание Rosemount 8600 Series
Страница 1: ...Reference Manual 00809 0100 4860 Rev BF May 2019 Rosemount 8600 Series Vortex Flow Meter ...
Страница 2: ......
Страница 8: ...viii Reference Manual 00809 0100 4860 Rev BF Title Page May 2019 Title Page ...
Страница 10: ...Reference Manual 00809 0100 4860 Rev BF Introduction May 2019 2 Rosemount 8600 Vortex Flow Meter ...
Страница 110: ...Reference Manual 00809 0100 4860 Rev BF Specifications and Reference Data May 2019 102 Rosemount 8600 Vortex Flow Meter ...
Страница 112: ...Reference Manual 00809 0100 4860 Rev BF Product Certifications May 2019 104 Rosemount 8600 Vortex Flow Meter ...
Страница 129: ......