3.3
Required tools and components
You will need the following tools and components to install the gas chromatograph (GC).
•
Zero grade carrier gas:
—
99.995% pure
—
Less than 5 ppm water
—
Less than 0.5 ppm hydrocarbons
•
High pressure dual-stage regulator for the carrier gas cylinder
—
High side up to 3,000 psig (207 barg)
—
Gauge (psig)
—
Low side capable of controlling pressure up to 150 psig (10 barg)
•
Calibration standard gas with correct number of components and concentrations
•
Dual-stage regulator for the calibration gas cylinder with a low pressure side capable of
controlling pressure up to 30 psig (2.07 barg)
•
Sample probe regulator (fixture for procuring the stream or sample gas for
chromatographic analysis)
•
Coalescing filter
•
Membrane filter
•
⅛
-in stainless steel tubing
—
For connecting calibration gas to the GC
—
For connecting carrier gas to the GC
—
For connecting stream gas to the GC
—
Sulfinert tubing required if sulfur components are present in calibration gas
•
Heat tracing, as required for sample transport and calibration lines
•
Miscellaneous tube fittings, tubing benders, and tubing cutter
•
14 American wire gauge (AWG) (18 metric wire gauge [MWG]) or larger electrical
wiring and conduit to provide 120 or 240 Vac, single phase, 50 to 60 Hz, from an
appropriate circuit breaker and power disconnect switch.
•
Digital volt-ohm meter with probe-type leads
•
Flow measuring device
•
Open-end wrenches sized
¼
-in, 5/16-in, 7/16-in,
½
-in, 9/16-in, and
⅝
-in.
•
Torque wrench
Related information
Reference Manual
Getting started
2-3-9000-744
June 2022
Rosemount 700XA
45
Содержание Rosemount 700XA
Страница 1: ...Reference Manual 2 3 9000 744 Rev L June 2022 Rosemount 700XA Gas Chromatograph ...
Страница 6: ...TxD TD or Sout Transmit data or signal out 6 ...
Страница 30: ...Overview Reference Manual June 2022 2 3 9000 744 30 Emerson com Rosemount ...
Страница 100: ...Installation and start up Reference Manual June 2022 2 3 9000 744 100 Emerson com Rosemount ...
Страница 117: ...Figure 5 10 LSIV exploded view Reference Manual Operation and maintenance 2 3 9000 744 June 2022 Rosemount 700XA 117 ...
Страница 182: ...Operation and maintenance Reference Manual June 2022 2 3 9000 744 182 Emerson com Rosemount ...
Страница 223: ...Figure A 19 Chromatogram menu Reference Manual Local operator interface LOI 2 3 9000 744 June 2022 Rosemount 700XA 223 ...
Страница 262: ...Local operator interface LOI Reference Manual June 2022 2 3 9000 744 262 Emerson com Rosemount ...
Страница 270: ...Micro flame photometric detector µFPD Reference Manual June 2022 2 3 9000 744 270 Emerson com Rosemount ...
Страница 278: ...Pre defined Modbus map files Reference Manual June 2022 2 3 9000 744 278 Emerson com Rosemount ...
Страница 280: ......
Страница 281: ......
Страница 282: ......
Страница 293: ......