![Emerson Rosemount 644 with HART Protocol Скачать руководство пользователя страница 75](http://html.mh-extra.com/html/emerson/rosemount-644-with-hart-protocol/rosemount-644-with-hart-protocol_reference-manual_104732075.webp)
67
Reference Manual
00809-0200-4728, Rev SA
Troubleshooting
July 2018
Troubleshooting
Figure 5-1. Trim
5.4.1
Application: Linear offset (single-point trim solution)
1. Connect sensor to transmitter. Place sensor in bath between range points.
2. Enter known bath temperature using the Field Communicator.
5.4.2
Application: Linear offset and slope correction (two-point trim)
1. Connect sensor to transmitter. Place sensor in bath at low range point.
2. Enter known bath temperature using the Field Communicator.
3. Repeat at high range point.
Use the following procedures to perform a sensor trim on the Rosemount 644 Transmitter:
Field Communicator
1. Connect the calibration device or sensor to the transmitter. (If using an active calibrator, see
calibrator and EMF compensation” on page 69
2. Connect the communicator to the transmitter loop.
From the
HOME
screen, enter the Fast Key sequence.
The communicator will ask “Are you using an active calibrator?”
a. Select
No
if a sensor is connected to the transmitter
b. Select
Yes
if using a calibration device. By selecting yes, the transmitter will switch into active
calibration mode (see “Active Calibrator and EMF Compensation”). This is critical if the calibrator
requires constant sensor current for calibration. If using a calibration device that can accept
pulsed current, select “No.”
Single-point trim
Two-point trim
Transmitter System Curve
Site-Standard Curve
Device Dashboard Fast Keys
3, 4, 4, 1
R
e
sist
anc
e
(ohms)
Temperature
R
e
sist
anc
e
(ohms)
Temperature
Содержание Rosemount 644 with HART Protocol
Страница 2: ......
Страница 8: ...vi Reference Manual 00809 0200 4728 Rev SA Title Page July 2018 Title Page ...
Страница 48: ...40 Reference Manual 00809 0200 4728 Rev SA Configuration July 2018 Configuration ...
Страница 62: ...54 Reference Manual 00809 0200 4728 Rev SA Installation July 2018 Installation ...
Страница 72: ...64 Reference Manual 00809 0200 4728 Rev SA Operation and Maintenance July 2018 Operation and Maintenance ...
Страница 96: ...Reference Data July 2018 Reference Manual 00809 0200 4728 Rev SA Specifications and Reference Data 88 ...
Страница 111: ......