MAINT_ACTIVE
The MAINT_ACTIVE parameter displays which of the alarms is active. Only the condition
with the highest priority will be displayed. This priority is not the same as the MAINT_PRI
parameter described above. This priority is hard coded within the device and is not user
configurable.
MAINT_ALM
An alarm indicating that the device needs maintenance soon. If the condition is ignored,
the device will eventually fail.
I.3.3
Advisory Alarms
An advisory alarm indicates informative conditions that do not have a direct impact on the
primary functions of the device. There are five parameters associated with
ADVISE_ALARMS, they are described below.
ADVISE_ENABLED
The ADVISE_ENABLED parameter contains a list of informative conditions that do not have
a direct impact on the primary functions of the device. Below is a list of the advisories:
1. Non-Volatile Writes Deferred
2. Electronics Warning - Transducer Block
3. Alerts Simulation
ADVISE_MASK
The ADVISE_MASK parameter will mask any of the failed conditions listed in
ADVISE_ENABLED. A bit on means the condition is masked out from alarming and will not
be reported.
ADVISE_PRI
ADVISE_PRI designates the alarming priority of the ADVISE_ALM, see
default is 0 and the recommended values are 1 or 2.
ADVISE_ACTIVE
The ADVISE_ACTIVE parameter displays which of the advisories is active. Only the advisory
with the highest priority will be displayed. This priority is not the same as the ADVISE_PRI
parameter described above. This priority is hard coded within the device and is not user
configurable.
ADVISE_ALM
ADVISE_ALM is an alarm indicating advisory alarms. These conditions do not have a direct
impact on the process or device integrity.
I.4
Alarm priority
Alarms are grouped into five levels of priority:
Resource Transducer Block
Reference Manual
April 2019
00809-0100-4530
452
Rosemount 5300 Level Transmitter
Содержание Rosemount 5300 Series
Страница 1: ...Reference Manual 00809 0100 4530 Rev FA April 2019 Rosemount 5300 Level Transmitter Guided Wave Radar ...
Страница 26: ...Transmitter overview Reference Manual April 2019 00809 0100 4530 26 Rosemount 5300 Level Transmitter ...
Страница 112: ...Electrical installation Reference Manual April 2019 00809 0100 4530 112 Rosemount 5300 Level Transmitter ...
Страница 158: ...Configuration Reference Manual April 2019 00809 0100 4530 158 Rosemount 5300 Level Transmitter ...
Страница 168: ...Operation Reference Manual April 2019 00809 0100 4530 168 Rosemount 5300 Level Transmitter ...
Страница 244: ...Service and troubleshooting Reference Manual April 2019 00809 0100 4530 244 Rosemount 5300 Level Transmitter ...
Страница 380: ...Product certifications Reference Manual April 2019 00809 0100 4530 380 Rosemount 5300 Level Transmitter ...
Страница 396: ...Configuration parameters Reference Manual April 2019 00809 0100 4530 396 Rosemount 5300 Level Transmitter ...
Страница 487: ...Reference Manual 00809 0100 4530 April 2019 Reference Manual 487 ...