Product dielectric range
Note
Applicable to “Liquid Product Level” and “Solid Product Level” measurement modes.
Enter the range of the dielectric constant (DC) for the product in the tank. The selected
range of dielectric constant is used for setting the automatically calculated amplitude
thresholds.
Upper product dielectric constant
Note
Applicable to “Interface Level with submerged” and “Product Level and Interface Level”
measurement modes.
The Upper Product Dielectric Constant (DC) should be entered as accurately as possible.
The dielectric constant of the upper product is essential for calculating the interface level
and the upper product thickness. In addition, this value is used for setting the
automatically calculated amplitude thresholds.
In case it is unknown, use the Dielectric Constant Chart or Upper Product Dielectric
Constant Calculator embedded in the configuration tools as help when configuring the
DC.
If the dielectric constant of the lower product is significantly smaller than the dielectric
constant of water, you may need to make special adjustments. The dielectric constant of
water is 80.
Related information
Interface measurements
Example 2: Interface peak not found
Use the dielectric constant chart in Rosemount Radar Master
The Dielectric Constant Chart lists the dielectric constant of a large number of products.
Procedure
1. Select
Setup
→
Tank
.
2. Select the
Environment
tab.
3. Select the
Dielectric Constant Chart
button.
Use the upper product dielectric constant calculator in Rosemount
Radar Master
The Upper Product Dielectric Constant Calculator lets you calculate the dielectric constant
of the upper product based on the following input:
•
Actual upper product thickness
•
The dielectric constant value stored in the transmitter
•
The upper product thickness presented by the transmitter
Reference Manual
Configuration parameters
00809-0100-4530
April 2019
Reference Manual
389
Содержание Rosemount 5300 Series
Страница 1: ...Reference Manual 00809 0100 4530 Rev FA April 2019 Rosemount 5300 Level Transmitter Guided Wave Radar ...
Страница 26: ...Transmitter overview Reference Manual April 2019 00809 0100 4530 26 Rosemount 5300 Level Transmitter ...
Страница 112: ...Electrical installation Reference Manual April 2019 00809 0100 4530 112 Rosemount 5300 Level Transmitter ...
Страница 158: ...Configuration Reference Manual April 2019 00809 0100 4530 158 Rosemount 5300 Level Transmitter ...
Страница 168: ...Operation Reference Manual April 2019 00809 0100 4530 168 Rosemount 5300 Level Transmitter ...
Страница 244: ...Service and troubleshooting Reference Manual April 2019 00809 0100 4530 244 Rosemount 5300 Level Transmitter ...
Страница 380: ...Product certifications Reference Manual April 2019 00809 0100 4530 380 Rosemount 5300 Level Transmitter ...
Страница 396: ...Configuration parameters Reference Manual April 2019 00809 0100 4530 396 Rosemount 5300 Level Transmitter ...
Страница 487: ...Reference Manual 00809 0100 4530 April 2019 Reference Manual 487 ...