AMS
™
Device Manager or a HART
®
Field Communicator and select
Overview
→
Zero Calibration
→
Start
menu or
Service Tools
→
Maintenance
→
Zero
Calibration
→
Start
.
15. Close your application connection (MeterLink, AMS
™
Device Manager or HART
®
Field Communicator).
16. If replacing other electronics, continue with the following procedures before
replacing the end caps and sealing the enclosure.
17. If you are not replacing other electronics, replace the end caps and security latches
(3 mm Allen wrench required). If required, install the security seals (see
Junction Box for remote mount electronics option
).
18. Apply power to the meter.
This completes the I.S. Barrier Board replacement procedure.
If you encounter problems with this procedure, see the Emerson Customer Support
contact information on the back cover of this manual.
3.9.5
Replace the Power Supply Board
Procedure
1. If replacing the Power Supply board, disable electrical power to the meter.
2. Refer to
for the tools required to complete this procedure.
3. Disconnect the Transmitter Electronics Enclosure security seals, loosen the end cap
security latches with a 3mm Allen wrench and remove both end caps.
4. Use a 1/8 inch (3 mm) flat head screw driver and remove the terminal blocks from
the Power Supply board, the CPU Module and the Optional I/O Module. See
for board locations and associated terminal blocks.
5. Use a #1 Standard Phillips head screw driver and remove the four Backplane board
screws from the enclosure standoffs.
Maintenance and Troubleshooting Manual
Meter maintenance
00809-0100-3812
April 2022
Rosemount 3812 Liquid Ultrasonic Flow Meter
99
Содержание Rosemount 3812
Страница 4: ...Contents Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 4 Emerson com Rosemount ...
Страница 12: ...Routine Maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 12 Emerson com Rosemount ...
Страница 36: ...Troubleshooting Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 36 Emerson com Rosemount ...
Страница 126: ...Meter maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 126 Emerson com Rosemount ...
Страница 130: ...Engineering drawings Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 130 Emerson com Rosemount ...