Table 4-9: Relay Modes
(continued)
Relay mode
Purpose of relay mode
Auxiliary
functions
Digital input 1+2
Relay is energized while Digital Input 1 (IN1) and 2 (IN2)
are both active.
Yes
Off
Relay is always de-energized.
No
Set Point
On/Off Point Control using PV (D800).
Yes
Alarm
Relay is allocated to alarm indication duty.
Yes
Hi or Lo Alarm
High alarm limit (using On/Off Point Control):
•
The On Point must be greater than the Off Point.
•
Relay energizes when the PV (D800) is greater than the
On Point.
•
Relay de-energizes when the PV (D800) is less than the
Off Point.
Yes
Low alarm limit (using On/Off Point Control):
•
The On Point must be less than the Off Point.
•
Relay energizes when the PV (D800) is less than the Off
Point.
•
Relay de-energizes when the PV (D800) is greater than
the On Point.
An “A” is shown by the RL1 icon status (on the Full PV
Display) when there is a high or low alarm. (The section
is not applicable).
Totalizer
(Rosemount 3491 only) Relay outputs totalizer pulses.
No
Totalizer 1
(Rosemount 3492/3493 only) Relay outputs totalizer 1
pulses.
No
Totalizer 2
(Rosemount 3492/3493 only) Relay outputs totalizer 2
pulses.
No
Fault
Indicate fault condition by de-energizing relay.
No
Custom
Custom relay operation.
No
PV limits
Relay energizes while control unit PV value (D800) is
within on/off point limits.
Yes
On
Relay is always energized.
No
Auxiliary functions for relay modes
Relay modes automatically enable and disable special control functions, special alarms
and pumped volume totalizing as shown in
and
Reference Manual
Getting started
00809-0100-4841
January 2021
Emerson.com/Rosemount
97
Содержание Rosemount 3490 Series
Страница 4: ...4 ...
Страница 8: ...Introduction Reference Manual January 2021 00809 0100 4841 8 Emerson com Rosemount ...
Страница 16: ...Control unit overview Reference Manual January 2021 00809 0100 4841 16 Emerson com Rosemount ...
Страница 36: ...Installation Reference Manual January 2021 00809 0100 4841 36 Emerson com Rosemount ...
Страница 134: ...Getting started Reference Manual January 2021 00809 0100 4841 134 Emerson com Rosemount ...
Страница 158: ...Specifications and reference data Reference Manual January 2021 00809 0100 4841 158 Emerson com Rosemount ...
Страница 184: ...Additional features Reference Manual January 2021 00809 0100 4841 184 Emerson com Rosemount ...
Страница 189: ...Reference Manual 00809 0100 4841 January 2021 Emerson com Rosemount 189 ...