Rosemount

3300HT/3300HTVP/3400HT/

3400HTVP/3500P/3500VP

PERpH-X High Performance pH/ORP Sensors

Instruction Manual

LIQ-MAN-PERpH-X

Rev. J

April 2017

Содержание Rosemount 3300HT

Страница 1: ...Rosemount 3300HT 3300HTVP 3400HT 3400HTVP 3500P 3500VP PERpH X High Performance pH ORP Sensors Instruction Manual LIQ MAN PERpH X Rev J April 2017 ...

Страница 2: ...hasgkas ...

Страница 3: ...e program and maintain the product When replacement parts are required ensure that qualified people use replacement parts specified by Emerson Unauthorized parts and procedures can affect the product s performance place the safe operation of your process at risk and VOID YOUR WARRANTY Third party substitutions may result in fire electrical hazards or improper operation Ensure that all equipment do...

Страница 4: ... The following list provides concerning all revisions of this document Rev Level Date Notes J 04 2017 Updated information with new Emerson Style Guidelines Updated Ordering Information Specifications and Wiring Diagrams Added Accessories Information EC Declaration of Conformity and FM Installation Drawings About This Document ...

Страница 5: ...artup and Calibration 3 1 Electrode Preparation 21 3 2 pH Sensor Calibration 21 3 3 Recommended pH Sensor Standardization 21 Section 4 Maintenance 4 1 Electrode Cleaning 23 4 2 Reference Junction Replacement and Sensor Electrolyte Recharge 23 Section 5 Accessories 5 1 Accessories 25 EC Declaration of Conformity 27 Intrisicallly Safe Sensor Installation Drawing FM 29 Instruction Manual Table of Con...

Страница 6: ...Table of Contents Instruction Manual April 2017 LIQ MAN PERpH X ii Table of Contents ...

Страница 7: ...g 2859 kPa abs CRN Rating 200 psig at room temperature Wetted Materials Titanium Ryton Teflon Glass and User Specified O ring Material EPDM Viton or Kalrez Reference Electrode Double junction with replaceable process side electrolyte and Teflon junction Temperature Sensor Pt 100 RTD Process Connections Must use 1 in compression process connector P N 23166 00 or 23166 01 Weight Shipping Weight 1 lb...

Страница 8: ...s Connections Must use 1 in compression process connector P N 23166 00 or 23166 01 Can be inserted through a ball valve Weight Shipping Weight 21 inch Sensor 2 lbs 3 lbs 0 9 kg 1 4 kg 36 inch Sensor 3 lbs 4 lbs 1 4 kg 1 8 kg Table 1 4 Rosemount 3500P 3500VP sensor specifications Measured Range pH range 0 to 14 pH ORP range 1500 mV to 1500 mV Operating Temperature 41 to 248 F 5 to 120 C Storage Tem...

Страница 9: ...x ia IIC T6 Ta 20 C to 60 C Nonincendive for use in Class I Division 2 Groups A B C and D Temperature Class T6 Ta 20 C to 60 C Suitable for use in Class II and III Division 2 Groups E F and G Temperature Class T6 Ta 20 C to 60 C Hazardous Classified Locations IS I II III 1 ABCDEFG T6 Ta 60 C 1400332 Entity I 0 AEx ia IIC T6 Ta 60 C 1400332 Entity NI I 2 ABCD T6 Ta 60 C S II III 2 EFG T6 Ta 60 C Pe...

Страница 10: ... GPHT Glass 12 ORP O ring Material 30 EPDM 31 Viton 32 Kalrez Typical Model Number 3300HT 10 30 Table 1 6 Rosemount 3300HTVP sensor ordering information Model Sensor type 3300HTVP pH ORP Sensor Measuring Electrode 10 pH GPHT Glass 12 ORP O ring Material 30 EPDM 31 Viton 32 Kalrez Preamplifier Option _ No selection 70 SMART Preamplifier 1 Typical Model Number 3300HTVP 10 30 70 1 Only available if s...

Страница 11: ...BNC 1 62 15 ft cable without BNC 2 Typical Model Number 3400HT 10 21 30 62 1 For use with sensor head junction boxes 2 For wiring directly to transmitter or junction box Table 1 7 Rosemount 3400HTVP sensor ordering information Model Sensor type 3400HT pH ORP Sensor Measuring Electrode 10 pH GPHT Glass 12 ORP Sensor Length 21 21 in Titanium Tube 22 36 in Titanium Tube O ring Material 30 EPDM 31 Vit...

Страница 12: ...Kalrez Typical Model Number 3500P HT 01 10 21 30 1 Preamplifier is SMART if selected with option 10 Preamplifier is standard if selected with option 12 Table 1 7 Rosemount 3500VP sensor ordering information Model Sensor type 3500VP pH ORP Sensor Electrolyte Selection BF Bio Film Resistant HT High Temperature MR Metal Resistant OR Oil Resistant PR Poisoning Resistant SR Scaling Resistant Preamplifi...

Страница 13: ...00HTVP below 23 F 5 C 3 Electrodes should be stored with a protective cap containing KCl solution 4 For overnight storage immerse the sensor in tap water or 4 pH buffer solution 5 A pH glass electrode has a limited shelf life of one year 2 2 Installation For sensor dimensions see Figure 2 1 to Figure 2 3 For sensor orientation and installation see Figure 2 4 to Figure 2 13 For wiring see Figure 2 ...

Страница 14: ...8 Installation Installation Instruction Manual April 2016 LIQ MAN PERpH X Figure 2 2 Rosemount 3300HT and 3300HTVP sensor dimensional drawing ...

Страница 15: ...Installation 9 Instruction Manual Installation LIQ MAN PERpH X April 2017 Figure 2 3 Rosemount 3400HT and 3400HTVP sensor dimensions with optional Ball Valve PN 23240 00 ...

Страница 16: ...10 Calibration and Maintenance Installation Instruction Manual April 2017 LIQ MAN PERpH X Figure 2 4 Rosemount 3400HT and 3400HTVP sensor dimensions with optional Ball Valve PN 23765 00 ...

Страница 17: ... X April 2017 Figure 2 5 Rosemount 3300HT and 3300HTVP flow through and insertion installation 1 in pipe Tee PN 2002011 with 1 in threaded connections Figure 2 6 Rosemount 3500P and 3500VP sensor dimensional drawing Note The sensor must be mounted at least 10 above the horizon ...

Страница 18: ... 3500VP Jet Spray Cleaner PN 12707 00 for Submersion Installation This accessory is especially useful for keeping the sensor clean in dirty ponds or tanks It can be mounted using the Handrail Mounting Assembly or a similar submersion accessory Figure 2 8 Rosemount 3500P and 3500VP flow through installation ...

Страница 19: ...truction Manual Installation LIQ MAN PERpH X April 2017 Figure 2 9 Rosemount 3500P or 3500VP for Submersion Installation using the Handrail Mounting Assembly PN 11275 01 All parts shown are supplied Sensor sold separately ...

Страница 20: ... and 3400HTVP mounting details retraction version Figure 2 11 Exploded view of 1 in ball valve kit PN 23240 00 used with process connector PN 23166 00 or PN 23166 01 Figure 2 12 Exploded view of 1 in ball valve kit PN 23765 00 Ball valve kit includes 11 2 x 1 reducer 11 2 close nipple and 11 2 ball valve ...

Страница 21: ...Installation 15 Instruction Manual Installation LIQ MAN PERpH X April 2017 Figure 2 13 Rosemount 3300HT or 3300HTVP typical submersion installation ...

Страница 22: ...sensor back completely until the retraction stop collar is reached 3 Close the ball valve slowly If there is resistance the valve may be hitting the sensor Double check that the sensor has been retracted to the retraction stop collar 4 The male connector body A may now be completely unthreaded from the reducing coupling and the sensor removed for servicing WARNING System pressure may cause the sen...

Страница 23: ... Gray Reference Shield Red RTD in White Red RTD Sense White RTD Return Orange mV pH in XMT 5VDC V SNSR RTD SENSE SNS RTD IN pH pH in 5VDC V SNSR RTD RTN RTN Ref SHLD SHLD GUARD pH SHLD SHLD GUARD REF REF IN GND SOL GND 12 1 4 5 10 11 3 6 9 7 2 8 1 4 5 10 3 6 9 7 2 8 Connect to instrument ground or as shown ID inner drain No Connection Rosemount intruments 56 1056 1057 1066 For 3300HT 3300HTVP 3400...

Страница 24: ...N RTN Ref SHLD SHLD GUARD pH SHLD SHLD GUARD REF REF IN GND SOL GND 12 1 4 5 10 11 3 6 9 7 2 8 1 4 5 10 3 6 9 7 2 8 White Red RTD Sense ID inner drain No Connection Rosemount intruments 56 1056 1057 1066 For Rosemount 3300HT 3300HTVP 3400HT 3400HTVP 3500P 02 3500VP 02 sensors Non preamp sensors Blue Cable PERpH X Sensors Preamplifier in Junction Box Remote Junction Box PN 23555 00 or Sensor Head J...

Страница 25: ...ange mV pH in XMT 12 1 4 5 10 11 3 6 9 7 2 8 1 4 5 10 3 6 9 7 2 8 Connect to instrument ground or as shown ID inner drain 5V 5VDC V SNSR RTD SENSE SNS RTD IN pH pH in 5VDC V SNSR RTD RTN RTN Ref SHLD SHLD GUARD pH SHLD SHLD GUARD REF REF IN GND SOL GND Rosemount intruments 56 1056 1057 1066 For 3300HTVP 70 3400HTVP 70 3500P 01 3500VP 01 sensors Preamplifier in sensor Blue Cable PERpH X Sensors Pre...

Страница 26: ...k mV pH Sheild Blue Solution Ground Gray Reference in White Gray No connection Red RTD in White RTD Return Inner Drain Black mV pH in XMT 12 1 4 5 10 11 3 6 9 7 2 8 1 4 5 10 3 6 9 7 2 8 12 11 ID inner drain Connect to instrument ground or as shown White Red RTD Sense 5VDC V SNSR RTD SENSE SNS RTD IN pH pH in 5VDC V SNSR RTD RTN RTN Ref SHLD SHLD GUARD pH SHLD SHLD GUARD REF REF IN GND SOL GND Brow...

Страница 27: ...d contact Allow sensor to equilibrate to the buffer temperature and wait for reading to stabilize Value of buffer can now be acknowledged by the transmitter 2 Once the first buffer has been acknowledged by the transmitter rinse the buffer solution off of the sensor with distilled or deionized water 3 Repeat steps 1 and 2 using the second buffer solution 4 The theoretical slope value according to t...

Страница 28: ...22 Startup and Calibration Starup and Calibration Instruction Manual April 2017 LIQ MAN PERpH X ...

Страница 29: ...ill gel to accomplish this procedure 1 Remove the junction cap by turning counter clockwise 2 Remove the liquid junction by pulling the junction straight out 3 Remove the old reference fill gel by rinsing with water 4 Fill the reference fill chamber with the reference fill gel using the syringe and remove any air bubbles Top off the reference fill chamber until it is completely filled 5 Replace th...

Страница 30: ...24 Maintenance Maintenance Instruction Manual April 2017 LIQ MAN PERpH X ...

Страница 31: ...able 11 conductor shielded prepped for use with junction box 9200273 Extension cable 11 conductor shielded unprepped for use with junction box 915240 03 PVC flow through Tee in NPT process connection 915240 04 PVC flow through Tee 1 in NPT process connection 915240 05 PVC flow through Tee 1 in NPT process connection 2002011 CPVC flow through Tee 1 1 2 NPT process connection 11275 01 Handrail mount...

Страница 32: ... 24238 02 Poisoning resistant porous teflon liquid junction Viton O rings 24238 03 Oil resistant porous teflon liquid junction Viton O rings 24238 04 Scaling resistant porous teflon liquid junction EPDM O rings 24238 05 Metals resistant porous teflon liquid junction EPDM O rings Refill Kits 4 9210392 High temperature refill kit 0 to 145 C 9210426 Bio film resistant refill kit 0 to 60 C 9210425 Poi...

Страница 33: ...EC Declaration of Conformity 27 Instruction Manual EC Declaration of Conformity LIQ MAN PERpH X April 2017 EC Declaration of Conformity Note Please see website for most recent Declaration ...

Страница 34: ...28 EC Declaration of Conformity EC Declaration of Conformity Instruction Manual April 2017 LIQ MAN PERpH X ...

Страница 35: ...S WITHOUT PREAMPS ARE SIMPLE APPARATUS ZON E ZONE 0 7 CONTROL EQUIPMENT CONNECTED TO THE ASSOCIATED APPARATUS MUST NOT USE OR GENERATE MORE THAN 250V RESISTANCE BETWEEN INTRINSICALLY SAFE GROUND AND EARTH GROUND MUST BE LESS THAN OR EQUAL TO 1 OHM ANY FM APPROVED TRANSMITTER FOR DIVISION 1 WITH INTRINSICALLY SAFE OUTPUT PARAMETERS THIS FM APPROVED DEVICE MUST BE INSTALLED PER ITS INSTALLATION DRAW...

Страница 36: ...companies All other marks are the property of their respective owners The contents of this publication are presented for information purposes only and while effort has been made to ensure their accuracy they are not to be construed as warranties or guarantees express or implied regarding the products or services described herein or their use or applicability All sales are governed by our terms and...