![Emerson Rosemount 3152 Скачать руководство пользователя страница 25](http://html.mh-extra.com/html/emerson/rosemount-3152/rosemount-3152_reference-manual_104708025.webp)
Reference Manual
00809-0100-4835 Rev BE
February 2018
Rosemount 3150 Series
Installation
16
Figure 2-5 – Typical Transmitter
Mounting Bracket Configuration,
Traditional Flange
(1) (2)
Carbon Steel
Panel Mount
Stainless Steel
Panel Mount
Stainless Steel
2-Inch Pipe Mount
(3)
Bracket
Options
3152
&
3153
3154
Carbon Steel Panel Mount Bracket is
not available for the 3154
NOTE: All dimensions are nominal in inches (millimeters)
(1) Transmitter and bracket orientation with respect to gravity will not impact qualification
(2) Transmitters can alternatively be mounted inside bracket (as shown below) or with process connection positioned adjacent to bracket
(not shown)
(3) A pipe-mount kit with three (3) U-bolts (not shown) is also available. The use of a third U-bolt is necessary to meet KTA seismic
requirements in installations where the KTA Airplane Crash (APC) value of 8g is applicable. Please consult the applicable Product
Data Sheet (PDS) for ordering information.