12
Reference Manual
00809-0700-4001, Rev AA
Configuration
May 2017
Configuration
Table 2-3. Device Dashboard Fast Key Sequence
2.7.2
Verifying configuration with AMS Device Manager
Right click on the device and select
Configuration Properties
from the menu. Navigate the tabs to
review the transmitter configuration data.
2.7.3
Verifying configuration with local operator interface
Press any configuration button to activate the LOI. Select
VIEW CONFIG
to review the below
parameters. Use the configuration buttons to navigate through the menu.
The parameters to be
reviewed prior to installation include:
2.7.4
Verifying process variables configuration
This section describes how to verify that the correct process variables are selected.
Verifying process variables with a Field Communicator
Verifying process variables with AMS Device Manager
1. Right click the device and select
Overview
from the menu.
2. Select the
All Variables
button to display the primary, secondary, tertiary and quaternary variables.
2.8
Basic setup of the transmitter
This section provides the necessary steps for basic setup of a pressure transmitter. When installing in DP
level or DP flow applications, refer to
“Configuring scaled variable” on page 19
for setup instructions.
2.8.1
Setting pressure units
The pressure unit command sets the unit of measure for the reported pressure.
From the
HOME
screen, enter the Fast Key sequences listed
Function
Fast Key sequence
HART 7
HART 5
Alarm and Saturation Levels
2, 2, 2, 5
2, 2, 2, 5
Damping
2, 2, 1, 1, 5
2, 2, 1, 1, 5
Primary Variable
2, 1, 1, 4, 1
2, 1, 1, 4, 1
Range Values
2, 1, 1, 4
2, 1, 1, 4
Tag
2, 2, 7, 1, 1
2, 2, 7, 1, 1
Transfer Function
2, 2, 1, 1, 6
2, 2, 1, 1, 6
Units
2, 2, 1, 1, 4
2, 2, 1, 1, 4
Tag
Primary Variable
Units
Range Values
Transfer Function
Damping
Alarm and Saturation Levels
From the
HOME
screen, enter the Fast Key sequence
Device Dashboard Fast Keys
3, 2, 1
Содержание Rosemount 3051P
Страница 2: ......
Страница 8: ...6 Reference Manual 00809 0700 4001 Rev AA Contents May 2017 Contents ...
Страница 12: ...2 Reference Manual 00809 0700 4001 Rev AA Introduction May 2017 Introduction ...
Страница 36: ...26 Reference Manual 00809 0700 4001 Rev AA Configuration May 2017 Configuration ...
Страница 66: ...56 Reference Manual 00809 0700 4001 Rev AA Operation and Maintenance May 2017 Operation and Maintenance ...
Страница 74: ...64 Reference Manual 00809 0700 4001 Rev AA Troubleshooting May 2017 Troubleshooting ...
Страница 98: ...Product Certifications May 2017 Reference Manual 00809 0700 4001 Rev AA Product Certifications 88 ...
Страница 109: ......