51
Reference Manual
00809-0100-2240, Rev DA
October 2017
Temperature sensor element positions
The temperature elements are numbered from the bottom of the tank and upwards. Enter the position
of each element, measured as the distance from the Zero Level (Dipping Datum Plate) to the
temperature element. If you use average temperature elements, enter the position of the terminating
level of each sensor element.
Figure 5-1. Temperature sensor element positions
Tank average temperature calculation
You can exclude certain spot elements from the average temperature calculation. This may be useful if,
for example, the temperature close to the surface or close to the bottom of the tank deviates
significantly from the temperature in the rest of the tank. This may also be accomplished by setting an
appropriate value for the Insert Distance parameter, see
.
Note
A faulty temperature sensor will affect the temperature calculation. For more information see
.
16
2
3
1
Zero Level
Содержание Rosemount 2240S
Страница 2: ......
Страница 10: ...viii Reference Manual 00809 0100 2240 Rev DA Title Page October 2017 Title Page ...
Страница 16: ...6 Reference Manual 00809 0100 2240 Rev DA Introduction October 2017 Introduction ...
Страница 28: ...18 Reference Manual 00809 0100 2240 Rev DA Overview October 2017 Overview ...
Страница 116: ...106 Reference Manual 00809 0100 2240 Rev DA Configuration Operation October 2017 Configuration Operation ...
Страница 150: ...140 Reference Manual 00809 0100 2240 Rev DA Service and Troubleshooting October 2017 Service and Troubleshooting ...
Страница 164: ...154 Product Certifications Product Certifications October 2017 Reference Manual 00809 0100 2240 Rev DA ...
Страница 192: ...182 Reference Manual 00809 0100 2240 Rev DA Index October 2017 Index ...
Страница 193: ......