6. Select the
Configure>Manual Setup
option.
7. Set the device to Out Of Service (OOS) mode by clicking the
Change
button.
8. Select the desired tab (Device, Temperature Sensor, Water Level Sensor, etc.) and configure the device.
9. The Device tab lets you configure display units. It also provides the option to write protect the
Rosemount 2240S.
10.In addition to the Device tab there are various tabs that give you access to options such as
temperature sensor configuration and water level sensor configuration.
11.When configuration is finished, click the
Send
button to store the current configuration in the device
database.
12.Press the
Change
button to set the device back to operating (Auto) mode.
13.Press the
Close
button to close the window.