Mechanical installation
3.3.1
Mounting the graphical display
The Rosemount 2230 Graphical Field Display is designed for mounting on a wall or a pipe.
Wall mounting with bracket
The Rosemount 2230 Graphical Field Display can be mounted on a wall by using the mounting kit
supplied by Emerson Automation Solutions/Rosemount Tank Gauging.
1. Mount the bracket on the wall by using
four M8 screws and flat washers.
Note
! Countersunk screws are not
suitable.
2. Attach the Rosemount 2230 display to the
bracket on the wall by sliding it from the
top downwards.
3. Secure the display to the bracket by
tightening the locking screw.
70 mm
70 mm
94 mm
Ø 9 mm
94 mm
Locking screw
Bracket