17
Reference Manual
00809-0100-4140, Rev AA
January 2017
c.
Scroll down (
) and then select
OUTPUT DELAY
(
).
d. Follow on-screen instructions and edit a new delay setting in the range 0 to 3600.
Entering a delay outside this range results in ERROR being displayed.
3.
If viewing the mode setting:
a.
Select
VIEW CONFIG
(
).
b. Scroll down (
) until
O DELAY
and the delay setting appears.
4.
Exit the menu system by either waiting one minute for the
EXIT MENU?
prompt, or
scrolling down menus to find and select
BACK TO MENU
and
EXIT MENU
.
Note
When messages about the loop appear, take appropriate safe action and select “
”.
See
“Configuration basics” on page 10
for further information.
2.8.3
Media density
The measured frequency of the fork, when immersed in process medium, can be affected
by liquid density variations. As a result, the dry-to-wet and wet-to-dry switching points are
different for all types and varieties of process medium (
).
To overcome this, accurate switching points can be calculated by the Rosemount 2140 and
Rosemount 2140:SIS after a suitable density band is selected for the process medium.
Media Density options the Rosemount 2140 and Rosemount 2140:SIS are:
0.4 – 0.6 SG
(for a 400 to 600 kg/m
3
range – e.g. propane)
0.5 – 0.9 SG
(for a 500 to 900 kg/m
3
range – e.g. alcohol)
0.8 – 1.3 SG
(for a 800 to 1300 kg/m
3
range – e.g. water)
(default selection)
1.2 – 3.0 SG
(for a 1200 to 3000 kg/m
3
range – e.g. acid)
Extra options on the Rosemount 2140 are:
Low compacted sediment
Medium compacted sediment
High compacted sediment
Extreme compacted sediment
Note
Versions of the Rosemount 2140 and Rosemount 2140:SIS with the Extended Features
Package (EFP) can use
Media Learn
) to automatically select a density band.
For all other versions, keep the default setting for Media Density when the liquid specific
gravity is unknown.
Fast keys
2, 2, 1, 1, 3
Содержание Rosemount 2140
Страница 1: ...Reference Manual 00809 0100 4140 Rev AA January 2017 Rosemount 2140 Level Detector Vibrating Fork ...
Страница 2: ......
Страница 8: ...VI Reference Manual 00809 0100 4140 Rev AA Contents January 2017 Contents ...
Страница 14: ...4 Reference Manual 00809 0100 4140 Rev AA Introduction January 2017 Introduction ...
Страница 112: ...102 Reference Manual 00809 0100 4140 Rev AA Troubleshooting January 2017 Troubleshooting ...
Страница 143: ......