Diagnostic message: failed - fix now
Table 6-2. Status: Failed – Fix Now
6.4.2
Diagnostic message: maintenance - fix soon
Table 6-3. Status: Maintenance – Fix Soon
Alert name
LOI screen
Problem
Recommended action
Electronics
Board Failure
FAIL
BOARD
A failure has been detected in
the electronics circuit board
1.Contact your local Emerson representative for
advice.
Critical
Electronics
Data Error
MEMORY
ERROR
A user written parameter does
not match the expected value
1.Confirm and correct all parameters listed in Device
Information.
2.Perform a Device Reset.
3.Replace electronics board.
Sensor
Failure
FAIL
SENSOR
A failure has been detected in
the fork sensor
1.Contact Customer Services.
Alert name
LOI screen
Problem
Recommended action
Electronics
Temperature
Beyond Limits
TEMP OUT
LIMITS
The temperature of the
electronics has exceeded its
safe operating range.
1.Confirm electronics temperature is within limits of
–40 to 175 °F (–40 to 80 °C) with or without the
LOI display.
2.Contact your local Emerson representative for
advice.
Power Advisory
Diagnostic
POWER
ADVISE
The level detector has
detected a deviation of the
terminal voltage outside of
configured limits. This may
indicate degraded electrical or
loop integrity
1.Check the dc power supply to ensure power is
correct, stable, and has minimal ripple.
2.Check loop wiring for degradation or improper
grounding.
3.Remove the wiring compartment cover
(considering hazardous location requirements) to
check for presence or water or corrosion.
If conditions have resumed to normal, clicking on
Reset Alert will clear the alert.
Electronics
Board
Parameter Error
MEMORY
WARN
(also in
advisory)
A device parameter does not
match the expected value.
The error does not affect level
detector operation or analog
output.
1.Contact your local Emerson representative for
advice.
Configuration
Buttons
Operator Error
STUCK
BUTTON
Device is not responding to
button presses.
1.Check configuration buttons are not stuck.
2.Contact your local Emerson representative for
advice.
Содержание Rosemount 2140
Страница 1: ...Reference Manual 00809 0100 4140 Rev AA January 2017 Rosemount 2140 Level Detector Vibrating Fork ...
Страница 2: ......
Страница 8: ...VI Reference Manual 00809 0100 4140 Rev AA Contents January 2017 Contents ...
Страница 14: ...4 Reference Manual 00809 0100 4140 Rev AA Introduction January 2017 Introduction ...
Страница 112: ...102 Reference Manual 00809 0100 4140 Rev AA Troubleshooting January 2017 Troubleshooting ...
Страница 143: ......