66
Reference Manual
00809-0100-4107, Rev BA
Section 5: Operation and Maintenance
April 2013
5.3
Recommended calibration tasks
Table 5-1. Basic and full calibration tasks
Note
For Rosemount 2051CA, Rosemount 2051TA range 0 and range 5 devices, an accurate absolute
pressure source is required.
5.4
Calibration overview
The Rosemount 2051 pressure transmitter is an accurate instrument that is fully calibrated in
the factory. Field calibration is provided to the user to meet plant requirements or industry
standards. Complete calibration of the Rosemount 2051 can be split into two halves: Sensor
Calibration and Analog Output Calibration.
Sensor Calibration allows the user to adjust the pressure (digital value) reported by the
transmitter to be equal to a pressure standard. The Sensor Calibration can adjust the pressure
offset to correct for mounting conditions or line pressure effects. This correction is
recommended. The calibration of the pressure range (pressure span or gain correction)
requires accurate pressure standards (sources) to provide a full calibration.
Like the Sensor Calibration, the analog output can be calibrated to match the user measurement
system. The Analog Output Trim (4-20mA/ 1-5 VOutput Trim) will calibrate the loop at the
4mA (1 V) and 20mA (5 V) points.
Absolute pressure transmitters (Rosemount 2051CA and Rosemount 2051TA) are
calibrated at the factory. Trimming adjusts the position of the factory characterization
curve. It is possible to degrade performance of the transmitter if any trim is done
improperly or with inaccurate equipment.
Field installation tasks
Bench calibration tasks
1.
Perform Sensor zero/lower Trim:
Compensate for mounting pressure
effects
a.
Refer to
for manifold
operation instructions to properly
drain/vent valves
2.
Set/check basic configuration
parameters
a.
Output units
b.
Range points
c.
Output type
d.
Damping Value
1.
Perform optional 4-20mA 1-5 Vdc
Output Trim
2.
Perform a Sensor Trim
a.
Zero/lower trim using line
pressure effect correction.
Reference
for
manifold drain/vent valve
operation instructions.
b.
Optional full scale trim. Sets the
span of the device and requires
accurate calibration equipment
c.
Set/check basic configuration
parameters.
Содержание Rosemount 2051
Страница 2: ......
Страница 4: ......
Страница 102: ...92 Reference Manual 00809 0100 4107 Rev BA Section 6 Troubleshooting April 2013 Troubleshooting ...
Страница 213: ...203 Reference Manual 00809 0100 4107 Rev BA Appendix B Product Certifications April 2013 Product Certifications ...
Страница 214: ...204 Reference Manual 00809 0100 4107 Rev BA Appendix B Product Certifications April 2013 Product Certifications ...
Страница 227: ......