13
Reference Manual
00809-0100-4107, Rev BA
Section 2: Configure
April 2013
Table 2-3. Rosemount 2051 Device Dashboard fast key sequence
2.5.2
Verifying configuration with AMS Device Manager
Right click on the device and select
Configuration Properties
from the menu. Navigate the
tabs to review the transmitter configuration data.
2.5.3
Verifying configuration with Local Operator Interface
Press any configuration button to activate the LOI. Select
VIEW CONFIG
to review the below
parameters. Use the configuration buttons to navigate through the menu.
The parameters to be
reviewed prior to installation include:
2.5.4
Verifying process variables configuration
This section describes how to verify that the correct process variables are selected.
Verifying process variables with a Field Communicator
Verifying process variables with AMS Device Manager
Right click on the device and select
Overview
from the menu.
1.
Click the
All Variables
button to display the primary, secondary, tertiary and quaternary
variables.
From the
HOME
screen, enter the fast key sequences listed
Function
Fast Key Sequence
HART 7
HART 5
Alarm and Saturation Levels
2, 2, 2, 5
2, 2, 2, 5
Damping
2, 2, 1, 1, 5
2, 2, 1, 1, 5
Primary Variable
2, 1, 1, 4, 1
2, 1, 1, 4, 1
Range Values
2, 1, 1, 4
2, 1, 1, 4
Tag
2, 2, 7, 1, 1
2, 2, 7, 1, 1
Transfer Function
2, 2, 1, 1, 6
2, 2, 1, 1, 6
Units
2, 2, 1, 1, 4
2, 2, 1, 1, 4
Tag
Primary Variable
Units
Range Values
Transfer Function
Damping
Alarm and Saturation
Levels
From the
HOME
screen, enter the fast key sequence
Device Dashboard Fast Keys
3, 2, 1
Содержание Rosemount 2051
Страница 2: ......
Страница 4: ......
Страница 102: ...92 Reference Manual 00809 0100 4107 Rev BA Section 6 Troubleshooting April 2013 Troubleshooting ...
Страница 213: ...203 Reference Manual 00809 0100 4107 Rev BA Appendix B Product Certifications April 2013 Product Certifications ...
Страница 214: ...204 Reference Manual 00809 0100 4107 Rev BA Appendix B Product Certifications April 2013 Product Certifications ...
Страница 227: ......