Advanced
Calibration offset
Difference between surface distance measured by transmitter and the same distance
measured by, for example, hand-dipping with a measurement tape. A positive Calibration
Offset value will increase the presented level value.
Figure C-6: Calibration Offset
A
C
D
B
A. Positive Calibration Offset value
B. Negative Calibration Offset value
C. Reported level
D. Actual level
Upper null zone
The Upper Null Zone defines how close to the transmitter's reference point a level value is
accepted. You can extend this value to block out disturbing echoes close to the antenna,
for example from the tank nozzle.
Note
Make sure the 20 mA value is below the Upper Null Zone. Measurements are not
performed within the Upper Null Zone (UNZ).
Related information
Reference Manual
Configuration parameters