Additional Communications Response Delay
is used to synchronize Modbus
communications with hosts that operate at a slower speed than the transmitter. The
value specified here will be added to each response the transmitter sends to the
host.
Tip
Do not set
Additional Communications Response Delay
unless required by your Modbus host.
6.2.2
Configure RS-485 terminals
ProLink III
Device Tools
>
Configuration
>
Communications
>
RS-485 Terminals
Overview
Modbus/RS-485 communication parameters control Modbus communication with the
transmitter's RS-485 terminals.
RS-485 terminal parameters include:
•
Baud rate
•
Parity
•
Stop bits
Procedure
1.
Set
Baud Rate
as appropriate for your network.
The following baud rates are supported:
•
4800
•
9600
•
19200
•
38400
2.
Set
Parity
as appropriate for your network.
The following parity bits are supported:
•
None
•
Odd
•
Even
3.
Set
Stop Bits
as appropriate for your network.
The following stop bits are supported:
•
1
•
2
Integrate the meter with the control system
48
Micro Motion
®
LNG Series Meters
Содержание Micro Motion LNG Series
Страница 1: ...Configuration and Use Manual MMI 20032809 Rev AA February 2017 Micro Motion LNG Series Meters ...
Страница 6: ...Contents iv Micro Motion LNG Series Meters ...
Страница 8: ...Getting Started 2 Micro Motion LNG Series Meters ...
Страница 12: ...Before you begin 6 Micro Motion LNG Series Meters ...
Страница 18: ...Quick start 12 Micro Motion LNG Series Meters ...
Страница 20: ...Configuration and commissioning 14 Micro Motion LNG Series Meters ...
Страница 50: ...Configure device options and preferences 44 Micro Motion LNG Series Meters ...
Страница 56: ...Integrate the meter with the control system 50 Micro Motion LNG Series Meters ...
Страница 58: ...Complete the configuration 52 Micro Motion LNG Series Meters ...
Страница 60: ...Operations maintenance and troubleshooting 54 Micro Motion LNG Series Meters ...
Страница 73: ...D1 and D2 density calibration using ProLink III Figure 9 1 Measurement support Configuration and Use Manual 67 ...
Страница 74: ...Measurement support 68 Micro Motion LNG Series Meters ...
Страница 94: ...Using ProLink III with the transmitter 88 Micro Motion LNG Series Meters ...
Страница 98: ...Default values and ranges 92 Micro Motion LNG Series Meters ...
Страница 99: ...Default values and ranges Configuration and Use Manual 93 ...