4.5.4
Configure
Density Cutoff
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Density
Overview
Density Cutoff
specifies the lowest density value that will be reported as measured. All
density values below this cutoff will be reported as 0.
Procedure
Set
Density Cutoff
to the value you want to use.
The default value for
Density Cutoff
is 0.2 g/cm
³
. The range is 0.0 g/cm
³
to 0.5 g/cm
³
.
Effect of
Density Cutoff
on volume measurement
Density Cutoff
affects liquid volume measurement. If the density value goes below
Density
Cutoff
, the volume flow rate is reported as 0.
Density Cutoff
does not affect gas standard
volume measurement. Gas standard volume values are always calculated from the value
configured for
Standard Gas Density
.
4.6
Configure temperature measurement
The temperature measurement parameters control how temperature data from the
sensor is reported.
4.6.1
Configure
Temperature Measurement Unit
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Temperature
Overview
Temperature Measurement Unit
specifies the unit that will be used for temperature
measurement.
Procedure
Set
Temperature Measurement Unit
to the option you want to use.
The default setting is
Degrees Celsius
.
Options for
Temperature Measurement Unit
The transmitter provides a standard set of units for
Temperature Measurement Unit
. Different
communications tools may use different labels for the units.
Configure process measurement
34
Micro Motion
®
LNG Series Meters
Содержание Micro Motion LNG Series
Страница 1: ...Configuration and Use Manual MMI 20032809 Rev AA February 2017 Micro Motion LNG Series Meters ...
Страница 6: ...Contents iv Micro Motion LNG Series Meters ...
Страница 8: ...Getting Started 2 Micro Motion LNG Series Meters ...
Страница 12: ...Before you begin 6 Micro Motion LNG Series Meters ...
Страница 18: ...Quick start 12 Micro Motion LNG Series Meters ...
Страница 20: ...Configuration and commissioning 14 Micro Motion LNG Series Meters ...
Страница 50: ...Configure device options and preferences 44 Micro Motion LNG Series Meters ...
Страница 56: ...Integrate the meter with the control system 50 Micro Motion LNG Series Meters ...
Страница 58: ...Complete the configuration 52 Micro Motion LNG Series Meters ...
Страница 60: ...Operations maintenance and troubleshooting 54 Micro Motion LNG Series Meters ...
Страница 73: ...D1 and D2 density calibration using ProLink III Figure 9 1 Measurement support Configuration and Use Manual 67 ...
Страница 74: ...Measurement support 68 Micro Motion LNG Series Meters ...
Страница 94: ...Using ProLink III with the transmitter 88 Micro Motion LNG Series Meters ...
Страница 98: ...Default values and ranges 92 Micro Motion LNG Series Meters ...
Страница 99: ...Default values and ranges Configuration and Use Manual 93 ...