Micro Motion

®

F-Series Sensor

Installation Manual

Installation Manual

P/N 20002298, Rev. A
April 2005

Содержание Micro Motion F Series

Страница 1: ...Micro Motion F Series Sensor Installation Manual Installation Manual P N 20002298 Rev A April 2005 ...

Страница 2: ...m Troubleshooting Refer to the transmitter manual for troubleshooting help Specifications Full product specifications can be found in the F Series Product Data Sheet which is available from the Micro Motion web site at www micromotion com Definitions The term MVD transmitter refers to the following transmitter models Models 1500 1700 2500 and 2700 Models 3500 and 3700 Installation steps Determinin...

Страница 3: ...e with the following electronics interfaces An integral core processor for connecting to a 4 wire remotely mounted transmitter or to a user supplied remote host Figure 1 An integrally mounted Model 1700 or 2700 transmitter Figure 2 A 9 wire junction box for connecting to a remotely mounted transmitter or a remotely mounted core processor Figure 3 An integrally mounted IFT9701 transmitter not avail...

Страница 4: ...in Figure 1 F Series sensor with core processor Approval tag Flow direction arrow Core processor housing Calibration tag Process connection Approval tag Calibration tag Core processor housing Purge fitting Purge fitting Sensor with purge fittings ...

Страница 5: ...tegrally mounted Model 1700 2700 transmitter Model 1700 or 2700 transmitter Approval tag Flow direction arrow Core processor housing Calibration tag Process connection Calibration tag Approval tag Purge fitting Purge fitting Model 1700 or 2700 transmitter Sensor with purge fittings ...

Страница 6: ...ore You Begin Figure 3 F Series sensor with junction box Junction box Approval tag Flow direction arrow Approval tag Process connection Calibration tag Purge fitting Purge fitting Junction box Sensor with purge fittings Calibration tag ...

Страница 7: ...eries sensor with integrally mounted IFT9701 transmitter Calibration tag Flow direction arrow Model IFT9701 transmitter Approval tag Process connection Model IFT9701 transmitter Approval tag Calibration tag Purge fitting Purge fitting Sensor with purge fittings ...

Страница 8: ... F300A sensor with flexible conduit extended core processor Flow direction arrow Core processor housing Calibration tag Process connection Flexible conduit Length 36 914 mm Minimum bend radius 2 1 8 54 mm Approval tag Purge fitting optional Purge fitting optional ...

Страница 9: ...xible conduit extended Model 1700 2700 transmitter Model 1700 or 2700 transmitter Flow direction arrow Core processor housing Calibration tag Process connection Approval tag Flexible conduit Length 36 914 mm Minimum bend radius 2 1 8 54 mm Purge fitting optional Purge fitting optional ...

Страница 10: ...7 Model F300A sensor with flexible conduit extended junction box Junction box Flow direction arrow Calibration tag Process connection Approval tag Flexible conduit Length 36 914 mm Minimum bend radius 2 1 8 54 mm Purge fitting optional Purge fitting optional ...

Страница 11: ...for the environment in which the sensor is installed See Figures 1 7 For installation in an area that requires intrinsic safety refer to the appropriate Micro Motion approval documentation shipped with the sensor or available from the Micro Motion web site at www micromotion com WARNING Failure to comply with requirements for intrinsic safety in a hazardous area could result in an explosion Make s...

Страница 12: ...ertical axis Figure 8 Maximum ambient temperature for F Series sensors with an integral core processor or Model 1700 2700 transmitter Hazardous area approvals may impose additional limits on ambient and process temperature For the ATEX T rating refer to the ATEX documentation shipped with the sensor or available on the Micro Motion web site at www micromotion com atex CSA ambient temperature limit...

Страница 13: ...toff valve downstream from the sensor is recommended to halt flow during the zeroing procedure For more information about zeroing refer to the instruction manual shipped with the transmitter Table 2 Maximum cable lengths Cable type Wire size Maximum length Micro Motion 9 wire to an MVD transmitter or core processor Not applicable 60 feet 20 meters Micro Motion 9 wire to all other transmitters Not ...

Страница 14: ...arrow The sensor has a flow direction arrow see Figures 1 7 to help you configure the transmitter for flow direction If possible install the sensor so that the flow direction arrow matches actual process flow Vertical pipeline If the sensor is installed in a vertical pipeline liquids and slurries should flow upward through the sensor Gases may flow upward or downward Liquids Tubes down Horizontal ...

Страница 15: ...F300A electronics The electronics of a Model F300A sensor are mounted at the end of a flexible conduit In addition to mounting the sensor housing in the pipeline as shown in Figure 10 the Model F300A electronics need to be mounted to a wall or instrument pole see Figure 11 A mounting bracket is shipped with the sensor for use in securing the electronics to a wall or instrument pole To mount the el...

Страница 16: ...ronics Mounting bracket wall mount Use 4 5 16 8 mm bolts to secure the bracket to a wall Slide the electronics into the bracket groove Mounting bracket instrument pole Use 2 5 16 8 mm U bolts to secure the bracket to an instrument pole Secure the electronics to the bracket with the 4 10 screws ...

Страница 17: ...cable see Junction box to a 9 wire remote transmitter or remote core processor on page 21 WARNING Failure to comply with the requirements for intrinsic safety in a hazardous area could result in an explosion Make sure the hazardous area specified on the sensor approval tag Figures 1 7 is suitable for the environment in which the sensor will be installed For installations in an area that requires i...

Страница 18: ...w The shielded heat shrink provides a shield termination suitable for use in the gland when using cable whose shield consists of foil and not a braid Proceed to step 2 of the wiring procedure below With armored cable where the shield consists of braid prepare the cable as described below but do not apply heat shrink Proceed to step 2 of the wiring procedure below 2 Remove the cover from the core p...

Страница 19: ... burning the cable apply heat 250 F or 120 C to shrink the tubing Figure 14 Applying the heat shrink g Position gland clamping insert so the interior end is flush with the heat shrink h Fold the cloth shield or braid and drain wires over the clamping insert and approximately 1 8 inch 3 mm past the O ring Figure 15 Folding the cloth shield i Install the gland body into the core processor housing co...

Страница 20: ...used for the RS 485 connection Connect the four wires to the numbered slots on the core processor Figure 17 Figure 17 Connecting the wires at the core processor 7 Reinstall and tighten the core processor housing cover 8 Additional wiring instructions for the transmitter can be found in the transmitter manual Note Never ground the 4 wire cable shield and shield drain wire s at the transmitter Core ...

Страница 21: ... can be grounded via the piping if the joints in the pipeline are ground bonded If the sensor is not grounded via the piping connect a ground wire to the internal or external grounding screw which is located on the core processor housing or junction box If national standards are not in effect follow these guidelines to ground the sensor Use copper wire 14 AWG 2 0 mm or larger wire size for groundi...

Страница 22: ...ss or set control devices for manual operation 2 Remove both purge plugs from the sensor case If purge lines are being used open the valve in the purge lines 3 Prepare the purge plugs for reinstallation by wrapping them with 3 5 turns of Teflon tape WARNING Removing a purge plug compromises the secondary containment of the sensor and could expose the user to process fluid The user should take all ...

Страница 23: ... process 6 The purge time is the amount of time required for full exchange of atmosphere to inert gas For each sensor size the purge time is different Refer to Table 3 If purge lines are being used increase the purge time to fill the additional volume of the purge line 7 Avoid pressurizing the sensor case At the appropriate time shut off the gas supply then immediately seal the purge outlet and in...

Страница 24: ...equipment that has not been removed from the original shipping package will be considered new and unused New and unused equipment requires a completed Return Materials Authorization form Used equipment All equipment that is not classified as new and unused is considered used This equipment must be completely decontaminated and cleaned before being returned Used equipment must be accompanied by a c...

Страница 25: ......

Страница 26: ...ocess Management Wiltonstraat 30 3905 KW Veenendaal The Netherlands T 31 0 318 495 670 F 31 0 318 495 689 Micro Motion Japan Emerson Process Management Shinagawa NF Bldg 5F 1 2 5 Higashi Shinagawa Shinagawa ku Tokyo 140 0002 Japan T 81 3 5769 6803 F 81 3 5769 6843 Micro Motion Asia Emerson Process Management 1 Pandan Crescent Singapore 128461 Republic of Singapore T 65 6777 8211 F 65 6770 8003 Mic...