Configure process measurement
Result:
• If the volume flow rate drops below 15 liters/second but not below 10 liters/second:
− The primary mA output will report zero flow.
− The frequency output will report the actual flow rate, and the actual flow rate will be used
in all internal processing.
• If the volume flow rate drops below 10 liters/second, both outputs will report zero flow, and 0 will
be used in all internal processing.
5.4
Configure gas standard volume flow measurement
Display
OFF-LINE MAINT→OFF-LINE CONFG→VOL→VOL TYPE GAS
ProLink II
ProLink→Configuration→Flow→Vol Flow Type
Field Communicator
Configure→Manual Setup→Measurements→Gas Standard Volume
The gas standard volume flow measurement parameters control how gas standard volume flow is
measured and reported.
The gas standard volume flow measurement parameters include:
•
Volume Flow Type
•
Standard Gas Density
•
Gas Standard Volume Flow Measurement Unit
•
Gas Standard Volume Flow Cutoff
Restriction
You cannot implement both liquid volume flow and gas standard volume flow. You must choose one or
the other.
5.4.1
Configure
Volume Flow Type
for gas applications
Display
OFF-LINE MAINT→OFF-LINE CONFG→VOL→VOL TYPE GAS
ProLink II
ProLink→Configuration→Flow→Vol Flow Type→Std Gas Volume
Field Communicator
Configure→Manual Setup→Measurements→Gas Standard Volume→Volume Flow Type→GSV
Volume Flow Type
controls whether liquid or gas standard volume flow measurement will be implemented.
Restriction
If you are using the petroleum measurement application, you must set
Volume Flow Type
to
Liquid
. Gas
standard volume measurement is incompatible with the petroleum measurement application.
74
Micro Motion 9739 MVD Transmitters
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...