Troubleshooting
Procedure
1.
Navigate to the drive gain data:
• If you are using ProLink II, choose
ProLink→Diagnostic Information
.
• If you are using a Field Communicator, choose
3 Service Tools→4 Maintenance→5 Diagnostic
Variables
2.
Observe and record drive gain data over an appropriate period of time, under a variety of process
conditions.
10.26
Check the pickoff voltage
Collecting pickoff voltage data
To know whether your pickoff voltage is unusually low, you must collect pickoff voltage data during the
problem condition and compare it to pickoff voltage data from a period of normal operation.
Table 10-10
Possible causes and recommended actions for low pickoff voltage
Possible cause
Recommended actions
Faulty wiring runs between the sensor and transmitter
Verify wiring between sensor and transmitter.
Process flow rate beyond the limits of the sensor
Verify that the process flow rate is not out of range of
the sensor.
Slug flow
Check for slug flow. See Section 10.24.
No tube vibration in sensor
•
Check for plugging.
•
Ensure sensor is free to vibrate (no mechanical
binding).
•
Verify wiring.
•
Test coils at sensor. See Section 10.27.1.
Moisture in the sensor electronics
Eliminate the moisture in the sensor electronics.
The sensor is damaged
Contact Micro Motion.
10.26.1
Collect pickoff voltage data
Prerequisites
You will need either ProLink II or a Field Communicator to collect the pickoff voltage data.
Configuration and Use Manual
197
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...