Troubleshooting
Procedure
1.
Set
HART Address
as appropriate for your HART network.
2.
Set
Loop Current Mode
to
Enabled
.
10.15
Check HART burst mode
Procedure
1.
Check to see if burst mode is enabled or disabled.
2.
If burst mode is enabled, disable it.
10.16
Check mA output trim
There are four mA output trim values: a 4 mA and a 20 mA trim value for each mA output.9739
MVD transmitters have a 4 and 20 mA trim value for each mA output. Verify the configuration of all
the trim values.
Procedure
Verify the trim values for the mA output.
Note
Micro Motion does not recommend attempting to trim an mA output if the reading is off by more than
±
200 microamps. If this is the case, contact Micro Motion Customer Service.
10.17
Check
Lower Range Value
and
Upper Range Value
Procedure
1.
Verify current process conditions.
2.
Check the configuration of the LRV and URV.
10.18
Check
mA Output Fault Action
Procedure
1.
Check the status alarms for fault conditions.
If no fault conditions are present, the source of the problem is something other than the mA
fault configuration.
2.
Check the configuration of
mA Output Fault Action
.
10.19
Check
Frequency Output Mode
Frequency Output Mode
is used only to defined the relationship between two frequency outputs. If your
transmitter is not configured for two frequency outputs,
Frequency Output Mode
is not causing your output
problem.
Configuration and Use Manual
193
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...