6
Wire the printer
Procedure
Wire the RS232-RS485 adapter terminals to the 5700 terminals.
Converter
Terminals
K2 ADE
•
Wire the K2 ADE terminal 1 to Channel E, RS-485B, terminal
10
•
Wire the K2 ADE terminal 2 to Channel E, RS-485A, terminal 9
K2
•
Wire the K2 terminal 3 to Channel E, RS-485B, terminal 10
•
Wire the K2 terminal 4 to Channel E, RS-485A, terminal 9
Figure 6-1: Wiring between the K2 terminals to the 5700 terminals
A. Transmitter with end-cap removed
Important
For a K2 ADE converter, use terminals 1 and 2 on the converter.
For more information, see RS485/RS232 adapter instructions supplied by the
manufacturer.
Installation Manual
Wire the printer
MMI-20027478
February 2020
Installation Manual
45