3.4
Channel power requirements
The supply voltage required by the 4200 transmitter depends on the total resistance in the
mA loop. This includes all sense resistance and wire resistance.
Channel A mA HART terminal requirements
Use the chart below to determine the required supply voltage for Channel A based on loop
resistance.
Figure 3-1: Channel A output supply voltage and loop resistance
A. Loop resistance (Ohms)
B. Supply voltage (Vs)
C. Maximum Loop Resistance (Ohms)
Table 3-2: For maximum loop resistance for Channel A
Condition
Equation
17.75V < Vs <19.6V
(Vs-17.75)/3.6mA
19.6V < Vs < 21.5V
(Vs-8.32V)/22mA
21.5V <Vs <30V
600 ohms
Note
If display backlighting is enabled (refer to
Configure the display backlight
input voltage required is 1V higher than the chart indicates.
Wiring the channels
Installation Manual
April 2019
MMI-20058013
24
Micro Motion
®
4200 2-Wire Transmitter
Содержание Micro Motion 4200
Страница 1: ...Installation Manual MMI 20058013 Rev AB April 2019 Micro Motion 4200 2 Wire Transmitter ...
Страница 4: ...Contents Installation Manual April 2019 MMI 20058013 4 Micro Motion 4200 2 Wire Transmitter ...
Страница 10: ...Planning Installation Manual April 2019 MMI 20058013 10 Micro Motion 4200 2 Wire Transmitter ...
Страница 32: ...Wiring the channels Installation Manual April 2019 MMI 20058013 32 Micro Motion 4200 2 Wire Transmitter ...
Страница 34: ...Power up the transmitter Installation Manual April 2019 MMI 20058013 34 Micro Motion 4200 2 Wire Transmitter ...