![Emerson Micro Motion 4200 Скачать руководство пользователя страница 12](http://html.mh-extra.com/html/emerson/micro-motion-4200/micro-motion-4200_configuration-and-use-manual_104632012.webp)
2.3 Commissioning wizards
The transmitter menu includes a
Guided Setup
to help you move quickly through the most common
configuration parameters. ProLink III also provides a commissioning wizard.
By default, when the transmitter starts up, the Guided Setup menu is offered. You can choose to use it or not.
You can also choose whether or not Guided Setup is displayed automatically.
•
To enter Guided Setup upon transmitter startup, choose
Yes
at the prompt.
•
To enter Guided Setup after transmitter startup, choose
Menu
>
Startup Tasks
.
•
To control the automatic display of Guided Setup, choose
Menu
>
Configuration
>
Guided Setup
.
For information on the ProLink III commissioning wizard, see the
Micro Motion ProLink III User Manual
.
As the commissioning wizards are self guided, they are not documented in detail.
2.4 Make a startup connection to the transmitter
For all configuration tools except the display, you must have an active connection to the transmitter to
configure the transmitter.
Procedure
Identify the connection type to use, and follow the instructions for that connection type in the appropriate
appendix.
Communications tool
Connection type to use
Instructions
ProLink III
HART
Using ProLink III with the transmitter
Field Communicator
HART
Using a Field Communicator with the
transmitter
2.5 Set the transmitter clock
Display
Menu
>
Configuration
>
Time/Date/Tag
ProLink III
Device Tools
>
Configuration
>
Transmitter Clock
Field Communicator
Configure
>
Manual Setup
>
Clock
The transmitter clock provides timestamp data for alerts, service logs, history logs, and all other timers and
dates in the system. You can set the clock for your local time or for any standard time you want to use.
Tip
You may find it convenient to set all of your transmitter clocks to the same time, even if the transmitters are
in different time zones.
Procedure
1. Select the time zone that you want to use.
2. If you need a custom time zone, select
Special Time Zone
and enter your time zone as a difference
from UTC (Coordinated Universal Time).
3. Set the time appropriately for the selected time zone.
Quick start
Configuration and Use Manual
April 2019
MMI-20048166
12
Micro Motion 4200 Transmitters
Содержание Micro Motion 4200
Страница 144: ...Measurement support Configuration and Use Manual April 2019 MMI 20048166 144 Micro Motion 4200 Transmitters ...
Страница 210: ...Using the transmitter display Configuration and Use Manual April 2019 MMI 20048166 210 Micro Motion 4200 Transmitters ...
Страница 218: ...Channel combinations Configuration and Use Manual April 2019 MMI 20048166 218 Micro Motion 4200 Transmitters ...