5.4.2
Configure Calculation Speed (Response Time)
Display
Not available
ProLink II
ProLink
>
Configuration
>
Device
>
Response Time
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Response
>
Calculation Speed
Field Communicator
Not available
Overview
Calculation Speed
is used to apply a different algorithm to the calculation of process variables
from the raw process data.
Calculation Speed
=
Special
produces faster and “noisier” response
to changes in the process.
In ProLink II,
Calculation Speed
is called
Response Time
.
Restriction
Calculation Speed
is available only on systems with the enhanced core processor.
Tip
You can use
Calculation Speed
=
Special
with either setting of
Update Rate
. The parameters control
different aspects of flowmeter processing.
Procedure
Set
Calculation Speed
as desired.
Option
Description
Normal
Transmitter calculates process variables at the standard speed.
Special
Transmitter calculates process variables at a faster speed.
5.5
Configure alarm handling
The alarm handling parameters control the transmitter’s response to process and device
conditions.
Alarm handling parameters include:
•
Fault Timeout
•
Status Alarm Severity
Configure device options and preferences
Configuration and Use Manual
89
Содержание MICRO MOTION 2700
Страница 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 142: ...Integrate the meter with the control system 134 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 154: ...Set up the Weights Measures application 146 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 156: ...Operations maintenance and troubleshooting 148 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 327: ...Device Tools Trending Figure C 24 Using ProLink III with the transmitter Configuration and Use Manual 319 ...
Страница 328: ...Using ProLink III with the transmitter 320 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 373: ......