4.7.3
Configure petroleum measurement using the
Field Communicator
1.
Choose
Online
>
Configure
>
Manual Setup
>
Measurements
>
Set Up Petroleum
.
2.
Specify the API table to use.
a. Open the Petroleum Measurement Source menu and select the API table
number.
Depending on your choice, you may be prompted to enter a reference
temperature or a thermal expansion coefficient.
b. Enter the API table letter.
These two parameters uniquely specify the API table.
3.
Determine how the transmitter will obtain temperature data for the petroleum
measurement calculations, and perform the required setup.
Option
Setup
Temperature data
from the sensor
a. Choose
Online
>
Configure
>
Manual Setup
>
Measurements
>
External
Pressure/Temperature
>
Temperature
.
b. Set
External Temperature
to
Disabled
.
A user-configured
static temperature
value
a. Choose
Online
>
Configure
>
Manual Setup
>
Measurements
>
External
Pressure/Temperature
>
Temperature
.
b. Set
External Temperature
to
Enabled
.
c. Set
Correction Temperature
to the value to be used.
Polling for tempera-
ture
a. Ensure that the primary mA output has been wired to support
HART polling.
b. Choose
Online
>
Configure
>
Manual Setup
>
Measurements
>
External
Pressure/Temperature
>
Temperature
.
c. Set
External Temperature
to
Enabled
.
d. Choose
External Polling
.
e. Set
Poll Control
to
Poll As Primary
or
Poll as Secondary
.
f. Determine whether you will use Polling Slot 1 or Polling Slot 2.
g. For the chosen slot, set
Ext Dev Tag
to the HART tag of the exter-
nal temperature device.
h. For the chosen slot, set
Polled Variable
to
Temperature
.
Tip
•
Poll as Primary
: No other HART masters will be on the network.
•
Poll as Secondary
: Other HART masters will be on the network.
The Field Communicator is not a HART master.
Configure process measurement
56
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Содержание MICRO MOTION 2700
Страница 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 142: ...Integrate the meter with the control system 134 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 154: ...Set up the Weights Measures application 146 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 156: ...Operations maintenance and troubleshooting 148 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 327: ...Device Tools Trending Figure C 24 Using ProLink III with the transmitter Configuration and Use Manual 319 ...
Страница 328: ...Using ProLink III with the transmitter 320 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 373: ......