4.7
Configure the petroleum measurement
application
The petroleum measurement application enables Correction for the effect of Temperature
on the volume of Liquids (CTL), by calculating and applying a Volume Correction Factor
(VCF) to volume measurement. Internal calculations are performed in compliance with
American Petroleum Institute (API) standards.
4.7.1
Configure petroleum measurement using ProLink II
1.
Choose
ProLink
>
Configuration
>
API Setup
.
2.
Specify the API table to use.
a. In
API Chapter 11.1 Table Type
, select the API table group.
b. In
Units
, select the the measurement units you want to use.
These two parameters uniquely specify the API table.
3.
If your API table is
53A
,
53B
,
53D
, or
54C
, set
Reference Temperature
to the appropriate
value for your application. Enter the value in °C.
4.
If your API table is
6C
,
24C
, or
54C
, set
Thermal Expansion Coefficient
to the appropriate
value for your application.
5.
Determine how the transmitter will obtain temperature data for the petroleum
measurement calculations, and perform the required setup.
Option
Setup
Temperature data
from the sensor
a. Choose
View
>
Preferences
.
b. Disable
Use External Temperature
.
A user-configured
static temperature
value
a. Choose
View
>
Preferences
.
b. Enable
Use External Temperature
.
c. Choose
ProLink
>
Configuration
>
Temperature
.
d. Set
External Temperature
to the value to be used.
Configure process measurement
Configuration and Use Manual
53
Содержание MICRO MOTION 2700
Страница 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 142: ...Integrate the meter with the control system 134 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 154: ...Set up the Weights Measures application 146 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 156: ...Operations maintenance and troubleshooting 148 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 327: ...Device Tools Trending Figure C 24 Using ProLink III with the transmitter Configuration and Use Manual 319 ...
Страница 328: ...Using ProLink III with the transmitter 320 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 373: ......