4.3.4
Configure Gas Standard Volume Flow Cutoff
Display
Not available
ProLink II
ProLink
>
Configuration
>
Flow
>
Std Gas Vol Flow Cutoff
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Flow
Field Communicator
Configure
>
Manual Setup
>
Measurements
>
GSV
>
GSV Cutoff
Overview
Gas Standard Volume Flow Cutoff
specifies the lowest gas standard volume flow rate that will
reported as measured. All gas standard volume flow rates below this cutoff will be
reported as 0.
Procedure
Set
Gas Standard Volume Flow Cutoff
to the value you want to use.
The default value for
Gas Standard Volume Flow Cutoff
is 0.0. The lower limit is 0.0. There is no
upper limit.
Interaction between Gas Standard Volume Flow Cutoff and
AO Cutoff
Gas Standard Volume Flow Cutoff
defines the lowest Gas Standard Volume flow value that the
transmitter will report as measured.
AO Cutoff
defines the lowest flow rate that will be
reported via the mA output. If
mA Output Process Variable
is set to
Gas Standard Volume Flow
Rate
, the volume flow rate reported via the mA output is controlled by the higher of the
two cutoff values.
Gas Standard Volume Flow Cutoff
affects both the gas standard volume flow values reported
via outputs and the gas standard volume flow values used in other transmitter behavior
(e.g., events defined on gas standard volume flow).
AO Cutoff
affects only flow values reported via the mA output.
Example: Cutoff interaction with AO Cutoff lower than Gas Standard Volume Flow
Cutoff
Configuration:
•
mA Output Process Variable
for the primary mA output:
Gas Standard Volume Flow Rate
•
Frequency Output Process Variable
:
Gas Standard Volume Flow Rate
•
AO Cutoff
for the primary mA output:
10
SLPM (standard liters per minute)
•
Gas Standard Volume Flow Cutoff
:
15
SLPM
Result: If the gas standard volume flow rate drops below 15 SLPM, the volume flow will be
reported as 0, and 0 will be used in all internal processing.
Configure process measurement
Configuration and Use Manual
39
Содержание MICRO MOTION 2700
Страница 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 142: ...Integrate the meter with the control system 134 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 154: ...Set up the Weights Measures application 146 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 156: ...Operations maintenance and troubleshooting 148 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 327: ...Device Tools Trending Figure C 24 Using ProLink III with the transmitter Configuration and Use Manual 319 ...
Страница 328: ...Using ProLink III with the transmitter 320 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 373: ......