a. 1 lb/sec = 16 oz/sec
b.
Mass Flow Conversion Factor
= 1/16 = 0.0625
4.
Set
Mass Flow Conversion Factor
to
0.0625
.
5.
Set
Mass Flow Label
to
oz/sec
.
6.
Set
Mass Total Label
to
oz
.
4.1.2
Configure Flow Damping
Display
Not available
ProLink II
ProLink
>
Configuration
>
Flow
>
Flow Damp
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Flow
Field Communicator
Configure
>
Manual Setup
>
Measurements
>
Flow
>
Flow Damping
Overview
Damping is used to smooth out small, rapid fluctuations in process measurement.
Damping
Value
specifies the time period (in seconds) over which the transmitter will spread changes
in the reported process variable. At the end of the interval, the reported process variable
will reflect 63% of the change in the actual measured value.
Procedure
Set
Flow Damping
to the value you want to use.
The default value is 0.8 seconds. The range depends on the core processor type and the
setting of
Update Rate
, as shown in the following table.
Core processor type
Update Rate
setting
Flow Damping
range
Standard
Normal
0 to 51.2 seconds
Special
0 to 10.24 seconds
Enhanced
Not applicable
0 to 51.2 seconds
Tips
•
A high damping value makes the process variable appear smoother because the reported value
changes slowly.
•
A low damping value makes the process variable appear more erratic because the reported value
changes more quickly.
•
The combination of a high damping value and rapid, large changes in flow rate can result in
increased measurement error.
•
Whenever the damping value is non-zero, the reported measurement will lag the actual
measurement because the reported value is being averaged over time.
Configure process measurement
26
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Содержание MICRO MOTION 2700
Страница 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 142: ...Integrate the meter with the control system 134 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 154: ...Set up the Weights Measures application 146 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 156: ...Operations maintenance and troubleshooting 148 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 327: ...Device Tools Trending Figure C 24 Using ProLink III with the transmitter Configuration and Use Manual 319 ...
Страница 328: ...Using ProLink III with the transmitter 320 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 373: ......