4.
To disable scheduled execution:
•
To disable execution of a single scheduled test, set
Hours Until Next Run
to 0.
•
To disable recurring execution, set
Hours Between Recurring Runs
to 0.
•
To disable all scheduled execution, click
Turn Off Schedule
.
Manage scheduled test execution using ProLink III
1.
Choose
Device Tools
>
Diagnostics
>
Meter Verification
>
Schedule Meter Verification
.
2.
To schedule a single test or the first test in recurring execution, specify a value for
Hours Until Next Run
.
3.
To schedule recurring execution, specify a value for
Hours Between Recurring Runs
.
4.
To disable scheduled execution:
•
To disable execution of a single scheduled test, set
Hours Until Next Run
to 0.
•
To disable recurring execution, set
Hours Between Recurring Runs
to 0.
•
To disable all scheduled execution, click
Disable Scheduled Execution
.
Manage scheduled test execution using the
Field Communicator
1.
Navigate to the Smart Meter Verification menu:
•
Overview
>
Shortcuts
>
Meter Verification
•
Service Tools
>
Maintenance
>
Routine Maintenance
>
Meter Verification
2.
Choose
Automatic Verification
.
3.
To schedule a single test or the first test in recurring execution, specify a value for
Hrs Until Next Run
.
4.
To schedule recurring execution, specify a value for
Set Recurring Hours
.
5.
To disable scheduled execution:
•
To disable execution of a single scheduled test, set
Hrs Until Next Run
to 0.
•
To disable recurring execution, set
Set Recurring Hours
to 0.
•
To disable all scheduled execution, choose
Turn Off Schedule
.
11.3
Zero the flowmeter
Zeroing the flowmeter establishes a baseline for process measurement by analyzing the
sensor's output when there is no flow through the sensor tubes.
Important
In most cases, the factory zero is more accurate than the field zero. Do not zero the flowmeter unless
one of the following is true:
Measurement support
186
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Содержание MICRO MOTION 2700
Страница 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 142: ...Integrate the meter with the control system 134 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 154: ...Set up the Weights Measures application 146 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 156: ...Operations maintenance and troubleshooting 148 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 327: ...Device Tools Trending Figure C 24 Using ProLink III with the transmitter Configuration and Use Manual 319 ...
Страница 328: ...Using ProLink III with the transmitter 320 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 373: ......