View test result data using ProLink II
1.
Choose
Tools
>
Meter Verification
>
Run Meter Verification
and click
View Previous Test Results
and Print Report
.
The chart shows test results for all tests stored in the ProLink II database.
2.
(Optional) Click
Next
to view and print a test report.
3.
(Optional) Click
Export Data to CSV File
to save the data to a file on your PC.
View test result data using ProLink III
1.
Choose
Device Tools
>
Diagnostics
>
Meter Verification
and click
Previous Test Results
.
The chart shows test results for all tests stored in the ProLink III database.
2.
(Optional) Click
Next
to view and print a test report.
3.
(Optional) Click
Export Data to CSV File
to save the data to a file on your PC.
View test result data using the Field Communicator
1.
Navigate to the Smart Meter Verification menu:
•
Overview
>
Shortcuts
>
Meter Verification
•
Service Tools
>
Maintenance
>
Routine Maintenance
>
Meter Verification
2.
(Optional) If the Field Communicator database is out of date, choose
Upload Results
Data from Device
.
3.
To view data from the most recent test, choose
Most Recent Test Results
.
4.
To view data for all tests in the Field Communicator database:
a. Press
Show Results Table
.
Data from the most recent test is displayed.
b. Press
OK
to scroll through data from previous tests.
c. To exit the results table, press
Abort
.
Smart Meter Verification results interpretation
When the meter verification test is completed, the result is reported as Pass, Fail, or Abort.
(Some tools report the Fail result as Caution instead.)
Pass
The test result is within the specification uncertainty limit. In other words, the
stiffness of the left and right pickoffs match the factory values plus or minus the
specification uncertain limit. If transmitter zero and configuration match factory
values, the sensor will meet factory specifications for flow and density
measurement. It is expected that meters will pass meter verification every time
the test is run.
Measurement support
182
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Содержание MICRO MOTION 2700
Страница 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 142: ...Integrate the meter with the control system 134 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 154: ...Set up the Weights Measures application 146 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 156: ...Operations maintenance and troubleshooting 148 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 327: ...Device Tools Trending Figure C 24 Using ProLink III with the transmitter Configuration and Use Manual 319 ...
Страница 328: ...Using ProLink III with the transmitter 320 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 373: ......