The controls are not accessible in any other way.
10.2.1
Switch between secured and unsecured mode using
ProLink II
Prerequisites
Before switching to unsecured mode, ensure that you will be able to switch back to
secured mode. Because switching to unsecured mode requires breaking the physical seal,
switching back to secured mode may require a site visit from a certified Weights &
Measures inspector and reinstallation of the physical seal.
Ensure that the wiring is in place for a service port connection. You may need to break the
seal and remove the clamp.
Procedure
•
To switch from secured mode to unsecured mode:
1. Make a service port connection to your transmitter.
2. Choose
Plug-ins
>
Enable/Disable Custody Transfer
.
3. Click
Disable Custody Transfer
.
•
To switch from unsecured mode to secured mode:
1. Make a service port connection to your transmitter.
2. Choose
Plug-ins
>
Enable/Disable Custody Transfer
3. Click
Enable Custody Transfer
.
10.2.2
Switch between secured and unsecured mode using
ProLink III
Prerequisites
Before switching to unsecured mode, ensure that you will be able to switch back to
secured mode. Because switching to unsecured mode requires breaking the physical seal,
switching back to secured mode may require a site visit from a certified Weights &
Measures inspector and reinstallation of the physical seal.
Ensure that the wiring is in place for a service port connection. You may need to break the
seal and remove the clamp.
Procedure
•
To switch from secured mode to unsecured mode:
1. Make a service port connection to your transmitter.
2. Choose
Device Tools
>
Configuration
>
Weights & Measures
.
3. Set
Software Security
to
Disabled
.
•
To switch from unsecured mode to secured mode:
Operate the transmitter with the Weights & Measures application
Configuration and Use Manual
169
Содержание MICRO MOTION 2700
Страница 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 142: ...Integrate the meter with the control system 134 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 154: ...Set up the Weights Measures application 146 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 156: ...Operations maintenance and troubleshooting 148 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 327: ...Device Tools Trending Figure C 24 Using ProLink III with the transmitter Configuration and Use Manual 319 ...
Страница 328: ...Using ProLink III with the transmitter 320 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 373: ......