Available actions when Approvals=
NTEP
(continued)
Table 10-5:
Function
Transmitter status
Unsecured
Secured
Outputs set to Last Meas-
ured Value
Allowed
Not allowed
External pressure and tem-
perature data
Retrieved by polling
Allowed
Allowed
Written by Modbus or
HART host
Allowed
Not allowed
Testing and trimming out-
puts
mA output loop test
Allowed
Not allowed
mA output trim
Allowed
Not allowed
Frequency output loop
test
Not allowed
Not allowed
Discrete output loop test Allowed
Not allowed
Totalizers
Resetting
Allowed. If any single
totalizer is reset, all
other totalizers are re-
set automatically.
May or may not be allowed, depend-
ing on configuration. Can be per-
formed only if flow rate is zero. If any
single totalizer is reset, all other total-
izers are reset automatically.
Starting
Not allowed
N/A
Stopping
N/A
Not allowed
Inventories
Resetting
Not allowed
Not allowed
Sensor simulation
Allowed
Not allowed
Available actions when Approvals=
OIML
Table 10-6:
Function
Transmitter status
Unsecured
Secured
Connecting from ProLink II, ProLink III, the
Field Communicator, or a host
All supported connec-
tion types are available.
Physical security prevents access to
the service port terminals.
Modbus/RS-485, HART/RS-485, and
HART/Bell 202 connections can be
used to read data from the transmit-
ter, but software security prevents
writing to the transmitter.
Configuration
Reading
Allowed
Allowed
Changing
Allowed
Not allowed
Calibration
Zero
Allowed
Not allowed
Density
Allowed
Not allowed
Temperature
Allowed
Not allowed
Operate the transmitter with the Weights & Measures application
Configuration and Use Manual
167
Содержание MICRO MOTION 2700
Страница 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 142: ...Integrate the meter with the control system 134 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 154: ...Set up the Weights Measures application 146 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 156: ...Operations maintenance and troubleshooting 148 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 327: ...Device Tools Trending Figure C 24 Using ProLink III with the transmitter Configuration and Use Manual 319 ...
Страница 328: ...Using ProLink III with the transmitter 320 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 373: ......