The display is able to show a maximum of eight characters, including the decimal point.
For all totalizer values configured as display variables, the position of the decimal point on
the display is fixed to the configured precision of the display variable.
When the totalizer reaches the largest value that can be displayed in these circumstances:
•
The digits roll over.
•
The decimal point does not move.
•
The number of digits on the display does not change.
•
The internal totalizer is reset to 0.
For example, the value 99999.999 rolls over to 00000.000.
Tip
If you are not sure whether or not the totalizer value on the display has rolled over, check the value
by using another method to read the current data.
Notes
•
The mass and volume totalizers do not necessarily roll over at the same time.
•
This feature does not apply to inventories. The display presents large inventory values in the
standard manner.
10.1.3
Effect of the Weights & Measures application on process
measurement and outputs
When the Weights & Measures application is installed, process measurement and
reporting is modified to prevent unauthorized use of the data. Specific changes are
controlled by whether the transmitter is secured or unsecured.
Transmitter outputs and process data when Approvals=
NTEP
Table 10-3:
Function
Transmitter status
Unsecured
Secured
Outputs
mA output behavior
If reporting a flow variable, mA output reports
zero flow. Otherwise, normal.
Normal
Frequency output behavior Inactive (produces no pulses), even under fault
conditions
Normal
Discrete output behavior
Normal
Normal
Process variables All flow rates
Reported as zero
Reported normally
Density
Reported normally
Reported normally
Temperature
Reported normally
Reported normally
Totalizers
Values
Do not increment or decrement
Increment normally
Inventories
Values
Do not increment or decrement
Increment normally
Operate the transmitter with the Weights & Measures application
Configuration and Use Manual
165
Содержание MICRO MOTION 2700
Страница 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 142: ...Integrate the meter with the control system 134 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 154: ...Set up the Weights Measures application 146 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 156: ...Operations maintenance and troubleshooting 148 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 327: ...Device Tools Trending Figure C 24 Using ProLink III with the transmitter Configuration and Use Manual 319 ...
Страница 328: ...Using ProLink III with the transmitter 320 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Страница 373: ......