Procedure
Verify the configuration of Frequency Output Mode.
12.23
Check Frequency Output Scaling Method
If Frequency Output Scaling Method is set incorrectly, the Frequency Output may report an
incorrect value.
1.
Verify the configuration of the Frequency Output.
2.
If you changed the setting of Frequency Output Scaling Method, check the settings
of all other Frequency Output parameters.
12.24
Check Frequency Output Fault Action
The Frequency Output Fault Action controls the behavior of the Frequency Output if the
transmitter encounters an internal fault condition. If the Frequency Output is reporting a
constant value, the transmitter may be in a fault condition.
1.
Check the status alerts for active fault conditions.
2.
If there are active fault conditions, the transmitter is performing correctly. If you
want to change its behavior, consider the following options:
•
Change the setting of Frequency Output Fault Action.
•
For the relevant status alerts, change the setting of Alert Severity to
Ignore
.
Restriction
For some status alerts, Alert Severity is not configurable.
3.
If there are no active fault conditions, continue troubleshooting.
12.25
Check Flow Direction
If Flow Direction is set inappropriately for your process, the transmitter may report flow
data that is not appropriate for your requirements.
The Flow Direction parameter interacts with actual flow direction to affect flow values,
flow totals and inventories, and output behavior. For the simplest operation, actual
process flow should match the flow arrow that is on the side of the sensor case.
Procedure
1.
Verify the actual direction of process flow through the sensor.
2.
Verify the configuration of Flow Direction.
Troubleshooting
Configuration and Use Manual
187
Содержание Micro Motion 2500
Страница 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 20: ...Configuration and commissioning 12 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 124: ...Complete the configuration 116 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 130: ...Set up the Weights Measures application 122 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 132: ...Operations maintenance and troubleshooting 124 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 140: ...Transmitter operation 132 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 216: ...Using ProLink III with the transmitter 208 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 226: ...Default values and ranges 218 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 237: ...NE 53 history Configuration and Use Manual 229 ...